TeleType Communications

MM sets the Brewer program to its teletype (terminal) mode, whereby the operator may communicate directly with the Brewer via the computer keyboard and a set of low level commands.
Exit from teletype mode by typing EXIT. See Appendix H for more explanation on the TeleType feature.

List of commands

CommandVariablesUsageDescription
A Repeats the command. Usually use in conjunction with another command move. (I.E. R,0,7,5:O:A) Press home to delete twice to go exit.
B,x
x refers to the following
0. Turns off all lamp
1. Turns on the mercury lamp
2. Turns on the standard lamp
B,2This command controls the lamps
M,m,s
m refers to a motor number
1. Zenith prism
2. Azimuth Tracker
3. Iris
4. Filterwheel 1
5. Filterwheel 2
6. Filterwheel 3
9. Micrometer 2 (Bottom)
10. Micrometer 1 (Top)
11. Slitmask 1
12. Slitmask 2
13. Zenith tracker
 
s refer to the number of steps to move
M,10,500Moves motor m to step s. Note: when moving backwards (negative number) it will set that as the new reference point. Click here more motor information.
N,ss refers to the number of stepsN,500(Used for MKIII only.) Moves both micrometer 1 and 2 to step s. Note: when moving backwards (negative number) it will set that as the new reference point. Click here more motor information.
O O
R,p1,p2,t
p# refers to the slit positions
t refers to number of times to repeat
R,0,5,7This command is used to measure the light intensity according to the variables setup. Often this command is combined with the O command to output the results. R,0,5,7:O
SL SL
  - Turn the SL bulb on
  - Points the zenith down
  - Open the iris
  - Set FW1 to position 1
  - Set FW2 to position 0
TStart Location2End Location
Start Location and End Location 
can be any of the following
1. MAX - Maximum micrometer position
2. OP - operating micrometer position
3. MIN - minimum micrometer position
TOP2MAXThis command will move the micrometer 1 to max/op/min position to max/op/min. It assumes the micrometer is already in the start location. This is use to help test the micrometer range.
BStart Location2End Location
Start Location and End Location 
can be any of the following
1. MAX - Maximum micrometer position
2. OP - operating micrometer position
3. MIN - minimum micrometer position
BOP2MAXThis command will move the micrometer 2 to max/op/min position to max/op/min. It assumes the micrometer is already in the start location. This is use to help test the micrometer range.
exit exitexit's mm and return to menu ('quit' will do the same thing)

For new electronics only

CommandVariablesUsageDescription
?<command> OR !<command>See Operation InformationA variety of operational commands/status output for the brewer. for more.
I,m
m refers to a motor number
1. Zenith prism
2. Azimuth Tracker
3. Iris
4. Filterwheel 1
5. Filterwheel 2
6. Filterwheel 3
9. Micrometer 2 (Bottom)
10. Micrometer 1 (Top)
11. Slitmask 1
12. Slitmask 2
13. Zenith tracker
I,9This commands re-initializes the specified motor
SAVESAVEStores the current set of RAM configuration parameters