TeleType Communications

MM sets the Brewer program to its teletype (terminal) mode, whereby the operator may communicate directly with the Brewer via the computer keyboard and a set of low level commands.
Exit from teletype mode by typing EXIT. See Appendix H for more explanation on the TeleType feature.

List of commands

CommandVariablesUsageDescription
A Repeats the command. Usually use in conjunction with another command move. (I.E. R,0,7,5:O:A) Press home to delete twice to go exit.
B,x
B,2This command controls the lamps
M,m,s
M,10,500Moves motor m to step s. Note: when moving backwards (negative number) it will set that as the new reference point. Click here more motor information.
O O
R,p1,p2,t
R,0,5,7This command is used to measure the light intensity according to the variables setup. Often this command is combined with the O command to output the results. R,0,5,7:O
B,x SL
TMAX TMAXMoves the top micrometer to the max position. This is to test to ensure the micrometer is able to move to the max range.
TMIN TMINMoves the top micrometer to the min position. This is to test to ensure the micrometer is able to move to the max range.
BMAX BMAXMoves the bottom micrometer to the max position. This is to test to ensure the micrometer is able to move to the min range.
BMIN BMINMoves the bottom micrometer to the min position. This is to test to ensure the micrometer is able to move to the min range.

For new electronics only

CommandVariablesUsageDescription
?<command> OR !<command>See Operation InformationA variety of operational commands/status output for the brewer. for more.
I,m
I,9This commands re-initializes the specified motor
SAVESAVEStores the current set of RAM configuration parameters