User Tools

Site Tools


hv

This is an old revision of the document!


HV

High Voltage Test (SB)

The HV command invokes a test used to determine the optimum high voltage setting for the photomultiplier. Once the instrument is properly setup (see Appendix F - jumper J2 on the high voltage module must be moved from Pins 2-3 to Pins 1-2 after the Brewer power is turned off, then the power returned to perform the test. After the test is performed the brewer power must again be turned off and the jumper J2 on the high voltage module must be returned from Pins 1-2 back to Pins 2-3, and then the power restored for normal operation). The HV command starts an automatic high voltage test and writes the results to HVJJJYY.III.

Operating PositionHV Test Position
Operating PositionHV Test Position

The HVSET command can then be used to set the High Voltage. Alternately, after the instrument is returned to normal operation, an adjustment can be made while monitoring the high voltage during multiple AP tests.

This test is normally used in the factory during the final set-up stages of manufacture, or in the field if it is suspected that the setting is not correct. The Standard Lamp is turned on, the iris is opened, FW#1 is put to position 1, and FW#2 is set to position 0.

See the HVSET command, and Appendix F for more details on the HV test.

To run a manual high voltage test for multiboard (old electronics) use HVB.

Suggested Values

DescriptionSuggested values
Photomultiplier serial Number:Enter the brewer serial numberIE: 046
Preame Voltage levelEnter the discriminator (PMT) voltage52mm PMT are typically around -30mV, where as 22mm PMT are usually around -10mV
Min VoltageWhere to start measuring the high voltage test fromFor MB, usually around -900~-1000v
Max VoltageWhere to end the high voltage test atFor MV, usually around -1700
IncrementIncrement per levelFor MB, 50, for SB, 25
High Voltage(V) Measured Voltage(V)
900-4.5
950-4.75
1000-5.00
1050-5.20
1100-5.50
1150-5.75
1200-6.00
1250-6.25
1300-6.50
1350-6.75
1400-7.00
1450-7.25
1500-7.50
1550-7.75
1600-8.00
1650-8.25
1700-8.50
1750-8.75
1800-9.00
hv.1700830000.txt.gz · Last modified: 2023/11/24 12:46 by reno