DSP

DSP is a command to perform a dispersion on the Brewer Spectrometer. This command cannot be ran together with other commands as one command string, and must be ran by itself.

For more information refer to the dispersion SOP.

Lamps

Mercury (Hg)

The lines that are scanned on a Single Spectrophotometer are the following,

LineStart SlitEnd SlitWavelength
1022893.60
2052967.28

The lines that are scanned on a Double Spectrophotometer are the following,

LineStart SlitEnd SlitWavelength
1012893.60
2032967.28
3053341.48

Cadmium (Cd)

The lines that are scanned on a Single Spectrophotometer are the following,

LineStart SlitEnd SlitWavelength
3053080.82
4053133.17
5153261.06

The lines that are scanned on a Double Spectrophotometer are the following,

LineStart SlitEnd SlitWavelength
4053080.82
5053133.17
6053261.06
7053403.65
8153499.95
9553611.51