Line 14: Line 14:
 |Dead time (sec)| Dead time the instrument is set to| |Dead time (sec)| Dead time the instrument is set to|
 |WL cal step number| Current cal step for the instrument | |WL cal step number| Current cal step for the instrument |
-|Unused| |+|Slitmask Motor Delay|Multi-board electronics - This is the slitmask delay between moving from one slit to the next. The typical value for this is 80\\ New Electronics - This sets the speed of the slitmask motor, typically and usually left at 14.|
 |Umkehr Offset| Azimuth offset during Umkehr measurements | |Umkehr Offset| Azimuth offset during Umkehr measurements |
 |ND filter 0| Filer correction from Neutral Density (ND) filter #0 | |ND filter 0| Filer correction from Neutral Density (ND) filter #0 |
Line 24: Line 24:
 |Zenith steps/rev| The number of steps to perform a complete rotation of the zenith | |Zenith steps/rev| The number of steps to perform a complete rotation of the zenith |
 |Brewer Type| Version of Brewer spectrophotometer | |Brewer Type| Version of Brewer spectrophotometer |
-|COM Port #| |+|COM Port #| Select the comport being used. **if using dosbox or similar dos emulator, this is not the actual comport used on the computer, but the comport specified in the config. Typically left as 1|
 |o3 Temp coef hg| | |o3 Temp coef hg| |
 |n2 Temp coef hg| | |n2 Temp coef hg| |
Line 42: Line 42:
 |NO2/O3 Mode Change| | |NO2/O3 Mode Change| |
 |Grating Slope| The relationship of movement between micrometer #1 and micrometer #2 to maximize counts| |Grating Slope| The relationship of movement between micrometer #1 and micrometer #2 to maximize counts|
-|Grating Intercept| |+|Grating Intercept|:::|
 |Micrometer Zero| | |Micrometer Zero| |
 |Iris Open Steps| The number of steps it takes to open the Iris | |Iris Open Steps| The number of steps it takes to open the Iris |
Line 54: Line 54:
 |Release date| Date that the ICF was created| |Release date| Date that the ICF was created|
 |EXTRAS| | |EXTRAS| |
-|AUTO HG| | +|AUTO HG|Allows the software to automatically insert HG as required when running a schedule with auto routine. Typically an HG is ran every 2 hours or a change in temperature of 2 degrees, which ever happens first. \\ 0 - Disable \\ 1 - Enable
-|FW2 oscil| |+|FW2 oscil|Allows filterwheel 2 to keep changing with no limits during DS measurements. This usually happens when the sun gets covered/uncovered by moving clouds.|
 |HG FW#2 Pos| Position of filter wheel #2 when using the Hg bulb | |HG FW#2 Pos| Position of filter wheel #2 when using the Hg bulb |
 |Low Signal limit| The number of counts at which the instrument will switch to a lower neutral density filter or skip the test during a direct sun measurement| |Low Signal limit| The number of counts at which the instrument will switch to a lower neutral density filter or skip the test during a direct sun measurement|
 |Upper Signal limit| The number of counts at which the instrument will increase the neutral density filter during a direct sun measurement| |Upper Signal limit| The number of counts at which the instrument will increase the neutral density filter during a direct sun measurement|
-|Use cubic DSP| |+|Use cubic DSP|Enable the use of cubic dispersion. \\ 0 - Disable \\ 1 - Enable|
 |Skip DS if too dim| Enable skipping a Direct Sun (DS) measurement if counts are too low  | |Skip DS if too dim| Enable skipping a Direct Sun (DS) measurement if counts are too low  |
 |O3ETC-R6| Ozone Extraterrestrial Constant (ETC) minus the R6 standard lamp ratio | |O3ETC-R6| Ozone Extraterrestrial Constant (ETC) minus the R6 standard lamp ratio |