====== QL ====== === Quick UV Lamp Scan === QL routine is used for performing UV calibrations on the Brewer Spectrometer. It is a test command that results in a UV scan being performed with a Lamp (rather than the sky) being the source of radiation.\\ The zenith prism is rotated to the UV dome, FW#2 is set to the "uv FW#2 Pos" recorded in the instrument constants file, FW#1 is set to the 3 position, the iris is opened, and the tracker is rotated to the sun (if the tracker is left powered during the test).\\ The operator is asked for lamp number, and lamp-diffuser separation (ECCC's Default setup uses 40cm), and the radiation intensity is measured in 1.5nm increments over the UVB range. Data is stored in a QLJJJYY.nnn file.\\ For more information, please refer to the {{ :docs:uv_sop_2016_v2.1.pdf |UV SOP}}. ===== TL;DR UV Calibration===== **Note**: Before using this section, please go through and understand how to perform the UV calibration via the {{ :docs:uv_sop_2016_v2.1.pdf |UV SOP}} - Run the pre-calibration command string. \\ For all instruments aside from MKIII:b1frhgsldtrsapFor MKIII instruments only:frhphgsldtrsap - With the calibration lamp on, check and make sure the zenith offset is correct via zetutututupzze2 - Run an RU or restart dosbox so it won't run out of string space in the middle. - Once the instrument is warmed up to temperature (Usually around 30c) run the uv calibration zehgb1qlhgzeb1qlhgsl or for MKIII instruments: zehphgb1qlhgzeb1qlhgsl - For subsequent bulbs you can run the following without waiting for the bulb warmup (wait time included) zew2hgb1qlhgzeb1qlhgsl For MKIII instruments: zew1hgb1qlhgzeb1qlhgsl