User Tools

Site Tools


singleboard

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revisionPrevious revision
Next revision
Previous revision
singleboard [2025/04/15 17:31] – [Troubleshooting] renosingleboard [2026/07/14 15:11] (current) – added heater information reno
Line 21: Line 21:
 {{ :docs:kz_splitboard_pcb_layout_b.pdf |}} {{ :docs:kz_splitboard_pcb_layout_b.pdf |}}
  
 +===== Heater Controller ===== 
 +The heater is controlled by a separate board that is normally attached to the back of the foreoptics support in front of the stock location of the heater. The minium temperature can be controlled via J1 
 +^Jumper setting^Minimum temperature^ 
 +|1-2|20°C| 
 +|2-3|10°C|
  
 ===== Mainboard Flat Cable Connections ===== ===== Mainboard Flat Cable Connections =====
Line 49: Line 53:
 {{:misc:20201019_074344_resized.jpg?300|}} {{:misc:20201019_074344_resized.jpg?300|}}
 |SL Voltage|HG Voltage|SL Current| |SL Voltage|HG Voltage|SL Current|
 +
 +===== Brewer Base Plate =====
 +<hidden Single Board Brewer Base Plate Drawing>{{:brewer_main:single_board_base_detail-1.jpg?400|}}
 +{{:brewer_main:single_board_base_holes-1.jpg?400|}}
 +{{:brewer_main:single_board_bm-c215_rev_1_spectrometer_weatherproof_base-1.jpg?400|}}</hidden>
 +\\
 ==== Troubleshooting ==== ==== Troubleshooting ====
 === AP Lamps Results === === AP Lamps Results ===
singleboard.1744738304.txt.gz · Last modified: 2025/04/15 17:31 by reno