| - | ===== HV ===== | + | ===== High Voltage Test (HV) ===== |
| - | === High Voltage | + | The high voltage test is used to determine the PMT response depending on the voltage provided to it. Upon entering the HV command, you will be prompted with the following. |
| + | < | ||
| + | Measure | ||
| + | Enter photomultiplier serial number: | ||
| + | Enter preamp voltage level: (default -30) | ||
| + | Enter min voltage: (default 1000) | ||
| + | Enter max voltage: (default 1600) | ||
| + | Enter increment: (default 50) | ||
| + | </ | ||
| + | The defaults may change from instrument to instrument, however those defaults values can be overridden by entering a different value and pressing enter. See [[hv# | ||
| - | The HV command invokes a test used to determine the optimum high voltage setting for the photomultiplier. | + | ==== High Voltage Test (SB¹) ==== |
| + | |||
| + | The HV command invokes a test used to determine the optimum high voltage setting for the photomultiplier. | ||
| ^Operating Position^HV Test Position^ | ^Operating Position^HV Test Position^ | ||
| |{{: | |{{: | ||
| Line 13: | Line 24: | ||
| See the [[hvset|HVSET]] command, and Appendix F for more details on the HV test. | See the [[hvset|HVSET]] command, and Appendix F for more details on the HV test. | ||
| - | To run a manual | + | ==== High Voltage Test (MB²) ==== |
| - | === Suggested Values === | + | - Ensure the instrument is able to run diagnostic data (As it still required to measure the SL Bulb) |
| + | - Turn the instrument off | ||
| + | - Take the lid off the Brewer. | ||
| + | - Take the cover off the Secondary power supply | ||
| + | - Turn the brewer on | ||
| + | - Run the command " | ||
| + | - Follow the dosbox prompt (See Suggested Values for more info) | ||
| + | - During the lamp warm up, set the multimeter to the voltage setting, and insert the leads to GND and HV on the panel beside the secondary power supply | ||
| + | - Once the lamp has warmed up and is waiting for the high voltage to be set, adjust the R202 pot until you are at the correct voltage - Refer to the table below for more info. Press the top left sighting button to when you have set the voltage.\\ {{: | ||
| + | - The Iris will move to the from the middle to the open position, then back to the middle, Once it is at the middle you may proceed to next voltage | ||
| + | - Repeat until you have completed the High voltage test | ||
| + | |||
| + | Note: Prior to version 7.33, to run a high voltage test for multiboard electronics, you will need to use the [[hvb|HVB]] | ||
| + | ==== Suggested Values | ||
| ^ ^Description^Suggested values^ | ^ ^Description^Suggested values^ | ||
| |Photomultiplier serial Number: | |Photomultiplier serial Number: | ||
| Line 43: | Line 67: | ||
| |1800|-9.00| | |1800|-9.00| | ||
| - | ==== Notes ==== | + | ===== Notes ===== |
| - | SB¹ - Indicates Single board electronics | + | SB¹ - Indicates Single board electronics\\ |
| MB² - Indicates Multi-board electronics | MB² - Indicates Multi-board electronics | ||
hv.1700830302.txt.gz · Last modified: 2023/11/24 12:51 by reno