User Tools

Site Tools


dvm

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revisionPrevious revision
Next revision
Previous revision
dvm [2022/09/28 20:34] renodvm [2023/06/19 17:31] (current) – Added instructions change the interface to rs232 reno
Line 7: Line 7:
 ^Serial Number^Asset Tag^Calibration Due Date^Location^ ^Serial Number^Asset Tag^Calibration Due Date^Location^
 |US34000577|J03995|January 12, 2020|To be sent for Calibration| |US34000577|J03995|January 12, 2020|To be sent for Calibration|
-|MY42002260|J81369|March 4, 2021|UV Cal room+|MY42002260|J81369|March 4, 2021|Spare - To be calibrated next
-|US36000699|  |March 9, 2021|?|+|US36000699|n/a|March 9, 2021|To be sent for calibration|
 |US36001261|  |March 3, 2021|?| |US36001261|  |March 3, 2021|?|
-|US36001852|  |January 202020To be sent for Calibration+|US36001852|???|October 272024|UV Cal room
-|US36001881| | |To be sent for calibration+|US36001881| |Oct 27, 2023|Remote 5
-|US36001878| | To be sent for calibration|+|US36001878| |Oct 27, 2023|Remote 4|
  
  
 +====== Setting dvm to RS232 mode ======
 +Because everytime it gets sent out the calibration, they don't set it back!!\\
 +Enter menu via <key>shift</key>+<key>< </key>\\
 +Use the <key>> </key> to get to F: I/O menu and press  to enter it\\
 +Press <key>> </key> once to get to interface and press <key>v</key> again\\
 +Press <key>> </key> to get to RS-232, press 
 +<key>auto/man</key>to save.
 +
 +The dvm will not be in RS232 mode.
  
dvm.1664397274.txt.gz · Last modified: 2022/09/28 20:34 by reno