dsp
Differences
This shows you the differences between two versions of the page.
| Both sides previous revisionPrevious revisionNext revision | Previous revision | ||
| dsp [2023/10/17 12:48] – andrew | dsp [2024/01/16 15:17] (current) – Added the notes section for additional information reno | ||
|---|---|---|---|
| Line 60: | Line 60: | ||
| - Press //y// or //n// and then press enter (n is pressed only if the lamp has been on for a minimum of 10 minutes). \\ {{: | - Press //y// or //n// and then press enter (n is pressed only if the lamp has been on for a minimum of 10 minutes). \\ {{: | ||
| - Press '// | - Press '// | ||
| - | - If the program determined wavelengths correspond with the table below, press '// | + | - If the program determined wavelengths correspond with the table below, press '// |
| - | + | ||
| - | ^Lamp Type^Line (LN)^Starting Slit^Ending SLit^Wavelenth AU^Brewer Model^ | + | |
| - | |Hg|1|0|1|2893.600|S, | + | |
| - | |Hg|2|0|3|2967.280|S, | + | |
| - | |Hg|3|0|5|3341.480|D| | + | |
| - | |Cd|4|0|5|3080.820|S, | + | |
| - | |Cd|5|0|5|3133.167|S, | + | |
| - | |Cd|6|0|5|3261.055|S, | + | |
| - | |Cd|7|0|5|3403.652|D| | + | |
| - | |Cd|8|1|5|3499.950|D| | + | |
| - | |Cd|9|5|5|3611.630|D| | + | |
| - | // | + | |
| - Press //enter// when ready. | - Press //enter// when ready. | ||
| - | - An example of the brewer screen while waiting for a dsp test to start. | ||
| - Typical screen display during a dsp test. Note: The highest counts should be near the middle of the scan, center portion of the center column. | - Typical screen display during a dsp test. Note: The highest counts should be near the middle of the scan, center portion of the center column. | ||
| - Once the dsp test has completed the brewer will return to menu mode. Type '// | - Once the dsp test has completed the brewer will return to menu mode. Type '// | ||
| Line 115: | Line 101: | ||
| + | ====== Notes ====== | ||
| + | Typically on instruments with a small slit 0 (and occasionally with instrument with a narrow offset slit 0) you may find the Fit to be closer to around 100pM. This is normal however it is suggested to use a quadratic dispersion as it usually is more accurate than a cubic dispersion.\\ | ||
| + | {{: | ||
| + | You can also see a difference in the slit 0 wavelength. This is not considered as a problem and just how the instrument was built. The dispersion will characterize the instrument properly\\ | ||
| + | {{:: | ||
dsp.1697546901.txt.gz · Last modified: 2023/10/17 12:48 by andrew
