disassembly
Differences
This shows you the differences between two versions of the page.
| Both sides previous revisionPrevious revision | |||
| disassembly [2025/12/09 19:10] – reno | disassembly [2025/12/18 12:21] (current) – reno | ||
|---|---|---|---|
| Line 1: | Line 1: | ||
| - | ====== PMT Removal ====== | + | ====== Disassembly |
| - | + | ||
| - | + | ||
| - | + | ||
| - | ===== Disassembly | + | |
| - | + | ||
| - | ==== Removing the Foroptics ==== | + | |
| - | + | ||
| - | - Remove the top 2 4-40 socket head screws and loosen 4-40 socket head screw on the side of the viewing window plate. {{: | + | |
| - | - Remove the viewing window plate | + | |
| - | - Remove the horseshoe clamp(3) by removing 2(1,2) 6-32 socket head screws.{{: | + | |
| - | - Remove the 3 slide lock DB9 connectors and 2 slide lock DB9 connectors on the side **Note: Pay attention to where these are as it will need to be connected to the right motors when the foreoptics is being re-installed. {{: | + | |
| - | - Carefully remove the foreoptics by lifting the foreoptics tube and sliding it to the left out of the mount.{{: | + | |
| - | + | ||
| - | When reinstalling the foreoptics, note that there is an alignment pin on the bottom of the foreoptics and the mount.\\ {{: | + | |
| - | ===== Removing | + | |
| - | - Remove the coax cable (1) | + | |
| - | - Remove the db9 connector (2) {{: | + | |
| - | - Remove the db9 connectors (1 & 2) | + | |
| - | - Carefully remove the thermister from the PMT (3) | + | |
| - | - Remove the 10-32 socket head screws (4,5,6,7) {{: | + | |
| - | - Carefully slide the entire PMT tube out (Note the orientation when you put the PMT back in later) | + | |
| - | - With the PMT standing up, remove the top cap (Note: It is a very fine thread, carefully remove the lid as there are electronics right under it.){{: | + | |
| - | - Remove the two 4-40 socket head screws {{: | + | |
| - | - Move the PCB board (High Speed Amp) out of the way and Loosen (do not remove) the 3 (1,2,3) 4-40 socket head screws (about 3 turns){{: | + | |
| - | - unscrew the threaded plate and carefully remove the PMT core {{:: | + | |
| - | - un-solder the resistor off the post | + | |
| - | - Tip 1: Due to the old solder, it maybe a good idea to add a little bit of new solder then remove it. | + | |
| - | - Tip 2: You don't need to completely remove all the solder, once most the solder is gone, use a pair of plyers to move the coiled up resistor off the post | + | |
| - | - Tip 3: there may be a bit of silicone on it, make sure you remove the silicone first {{: | + | |
| - | - Put on a pair of gloves! | + | |
| - | - Remove the PMT's shield to expose the PMT core{{:: | + | |
| - | - Slowly wiggle out the old PMT core {{: | + | |
| - | + | ||
| - | Notes for reinstalling the PMT | + | |
| - | - When reinstalling the " | + | |
| - | - Once the PMT core and PMT shield is reinstalled, | + | |
| - | - All the threads on the PMT are fine threads, be very careful and not cross thread anything | + | |
| - | - Repeat the steps backwards to reinstall everything. | + | |
| - | - In step 10, you do not need to thread it until it is tight, the 3 screws will keep it tight from letting it spin. Spin the thread so that the coax cable and db9 is in the correct orientation. | + | |
| - | - After re-installation, | + | |
| + | ===== Removing The PMT Core ===== | ||
| + | - Remove the foreoptics ([[foreoptics# | ||
| + | - Remove remove the PMT and the PMT core ([[pmt# | ||
| + | ===== Removing the Spectrometer ===== | ||
| + | [[spectrometer# | ||