Critical warnings from b-files' comments for days between 2025058 and 2025058 (0 sec, 2 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
hg:Lamp not on... hg test terminated F1= 0 F4= 59201B05825.254TORON#25413:27:381-1
hg:Lamp not detected f1= 0 f4= 59201B05825.254TORON#25413:27:391-1

Instrument constant changes for the week before 2025058 (4 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 5 rows)

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
2542025058117.7

Non-critical warnings from b-files' comments for days between 2025058 and 2025058 (0 sec, 4 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
ds:HOME key pressed199B05825.254TORON#25413:10:021-1
pd:Going to menu199B05825.254TORON#25413:10:071-1
ds:Suppressed oscillation of filter wheel.15B05825.254TORON#25418:56:261-1
ds:DS intensity too high ( 261935/cy) quitting.199B05825.254TORON#25418:57:091-1

Measurements and tests summary.


_jday_loc_serial_skcdsdtfzhghprssluvzs
058TORONTO#254ecsc,Next schedule will be ecsc (now ecsc ZS= 93 ZS= 104.543),Next schedule will be ecsc (now ecsc ZS= 93 ZS= 111.708),Next schedule will be ecsc (now ecsc ZS= 93 ZS= 118.703),Next schedule will be ecsc (now ecsc ZS= 93 ZS= 125.379),Next schedule will be ecsc (now ecsc ZS= 93 ZS= 131.529),Next schedule will be ecsc (now ecsc ZS= 93 ZS= 136.866),Next schedule will be ecsc (now ecsc ZS= 93 ZS= 141.007),Next schedule will be ecsc (now ecsc ZS= 93 ZS= 143.508),Next schedule will be ecsc (now ecsc ZS= 93 ZS= 93.74401),Next schedule will be ecsc (now ecsc ZS= 93 ZS= 99.136),Saving backup OP,Starting schedule ecsc22782922103
(0 sec)

Number of days since last dome/window cleaning (0 sec, 1 rows)

Missing files (0 sec, 1 rows)

Seriallast datadays ago
25420250590-1

Run/Stop results for last week. (0 sec, 1 rows)

SerialTotalFailed 0Failed darkFailed 1Failed 2Failed 3Failed 4Failed 5Failed double
254201000000

AP results for SB Brewers. (0 sec, 1 rows)

SerialJdayt1 offt1 ont2 offt2 ont3 offt3 onHV offHV on+12V off+12V on+5V off+5V on-12V off-12V on+24V off+24V on+5V sps off+5V sps onSL A offSL A onSL V offSL V onWatts
25420250587.47.57.87.87.27.01390.41390.412.112.15.15.1-12.0-12.024.224.25.05.00.01.50.010.1315.7

Standard lamp intensity analysis (0 sec, 1 rows)

Serialminmaxfilterfixable30-day history
#254258193025873200yes