Database last updated on Mon, 31 Aug 2026 22:41:07 +0000 by localhost
Critical warnings from b-files' comments for days between 2024160 and 2024160 (0 sec, 3 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| rl: | 2024 159 05 32 39 New value set for watermark SUPPLY.VOLTAGE.HIGH[+5VSEC]: 6.031851 Volts. | 9 | 3 | B16024.254 | TORON | #254 | 05:26:25 | | 1 | -1 |
| rl: | 2024 159 05 53 24 New value set for watermark LOOP.RUN.TIME: 653.999996 millisec. | 7 | 3 | B16024.254 | TORON | #254 | 05:26:31 | | 1 | -1 |
| rl: | 2024 160 11 10 03 PMT counter failure | 1 | 1 | B16024.254 | TORON | #254 | 05:26:44 | | 1 | -1 |
Instrument constant changes for the week before 2024160 (3 sec, 0 rows)
Non-critical warnings from b-files' comments for days between 2024160 and 2024160 (0 sec, 5 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| rl: | 2024 159 05 53 24 New value set for watermark LOOP.RUN.TIME: 653.999996 millisec. | 4 | 3 | B16024.254 | TORON | #254 | 05:26:31 | | 1 | -1 |
| rl: | 2024 160 07 34 39 New value set for watermark LOOP.RUN.TIME: 715.000010 millisec. | 3 | 3 | B16024.254 | TORON | #254 | 05:26:40 | | 1 | -1 |
| rl: | 2024 160 07 34 39 New value set for watermark LOOP.RUN.TIME: 715.000010 millisec. | 3 | 3 | B16024.254 | TORON | #254 | 05:26:40 | | 1 | -1 |
| fz: | Suppressed oscillation of filter wheel. | 4 | 5 | B16024.254 | TORON | #254 | 11:08:57 | | 1 | -1 |
| zs: | ZS is not done when solar zenith angle less than 30 degrees. Skipping. | 1 | 10 | B16024.254 | TORON | #254 | 15:43:17 | | 1 | -1 |
Measurements and tests summary.
| _jday | _loc | _serial | _skc | ds | dt | ed | fz | hg | hp | rs | sl | uv | zs |
| 160 | TORONTO | #254 | ecsc,Next schedule will be ecsc (now ecsc ZS= 93 ZS= 93.095) | 92 | 2 | 2 | 4 | 21 | 21 | 2 | 6 | 33 | 32 |
(0 sec)
Number of days since last dome/window cleaning (0 sec, 1 rows)
Run/Stop results for last week. (0 sec, 1 rows)
| Serial | Total | Failed 0 | Failed dark | Failed 1 | Failed 2 | Failed 3 | Failed 4 | Failed 5 | Failed double |
| 254 | 16 | 0 | 3 | 0 | 0 | 0 | 0 | 0 | 0 |
AP results for SB Brewers. (0 sec, 1 rows)
| Serial | Jday | t1 off | t1 on | t2 off | t2 on | t3 off | t3 on | HV off | HV on | +12V off | +12V on | +5V off | +5V on | -12V off | -12V on | +24V off | +24V on | +5V sps off | +5V sps on | SL A off | SL A on | SL V off | SL V on | Watts |
| 254 | 2024160 | 22.4 | 22.5 | 22.8 | 22.7 | 21.7 | 21.7 | 1389.5 | 1389.3 | 12.0 | 12.0 | 5.0 | 5.0 | -11.9 | -11.9 | 24.3 | 24.3 | 5.0 | 5.0 | 0.0 | 1.5 | 0.0 | 10.08 | 15.6 |
Standard lamp intensity analysis (0 sec, 1 rows)
| Serial | min | max | filter | fixable | 30-day history |
| #254 | 2564300 | 2566400 | 0 | yes |