Critical warnings from b-files' comments for days between 2024131 and 2024131 (0 sec, 3 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| rl: | 2024 130 05 37 10 New value set for watermark SUPPLY.VOLTAGE.HIGH[+5VSEC]: 6.037956 Volts. | 11 | 3 | | | #254 | 05:24:41 | | 1 | -1 |
| rl: | 2024 130 05 51 45 PMT counter failure | 1 | 1 | | | #254 | 05:24:46 | | 1 | -1 |
| rl: | 2024 130 05 55 07 New value set for watermark LOOP.RUN.TIME: 688.000011 millisec. | 9 | 3 | | | #254 | 05:24:48 | | 1 | -1 |
Instrument constant changes for the week before 2024131 (2 sec, 0 rows)
Non-critical warnings from b-files' comments for days between 2024131 and 2024131 (0 sec, 2 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| rl: | 2024 130 05 55 07 New value set for watermark LOOP.RUN.TIME: 688.000011 millisec. | 3 | 3 | | | #254 | 05:24:48 | | 1 | -1 |
| ds: | DS intensity too high ( 224465/cy) quitting. | 2 | 99 | | | #254 | 19:13:04 | | 1 | -1 |
Measurements and tests summary.
| _jday | _loc | _serial | _skc | ds | dt | ed | fz | hg | hp | rs | sl | uv | zs |
| 131 | TORONTO | #254 | ecsc,Next schedule will be ecsc (now ecsc ZS= 93 ZS= 94.149) | 78 | 2 | 2 | 4 | 16 | 17 | 2 | 7 | 30 | 42 |
(0 sec)
Number of days since last dome/window cleaning (0 sec, 1 rows)
Missing files (0 sec, 1 rows)
| Serial | last data | days ago |
| 254 | 2024128 | 003 |
Questionable Run/Stop results. (0 sec, 1 rows)
| Serial | Jday | slit 0 | dark rs | dark counts | slit 1 | slit 2 | slit 3 | slit 4 | slit 5 | slits 3+5 |
| 254 | 2024131 | 0.993 | 2.000 | 2 / 0 (1.7M) | 1.000 | 1.000 | 1.000 | 1.000 | 0.999 | 1.000 |
Run/Stop results for last week. (0 sec, 1 rows)
| Serial | Total | Failed 0 | Failed dark | Failed 1 | Failed 2 | Failed 3 | Failed 4 | Failed 5 | Failed double |
| 254 | 16 | 0 | 1 | 0 | 0 | 0 | 0 | 0 | 0 |
AP results for SB Brewers. (0 sec, 1 rows)
| Serial | Jday | t1 off | t1 on | t2 off | t2 on | t3 off | t3 on | HV off | HV on | +12V off | +12V on | +5V off | +5V on | -12V off | -12V on | +24V off | +24V on | +5V sps off | +5V sps on | SL A off | SL A on | SL V off | SL V on | Watts |
| 254 | 2024131 | 20.4 | 20.2 | 20.2 | 20.1 | 19.6 | 19.5 | 1388.8 | 1389.2 | 12.0 | 12.0 | 5.0 | 5.0 | -11.9 | -11.9 | 24.3 | 24.3 | 5.0 | 5.0 | 0.0 | 1.5 | 0.0 | 10.07 | 15.6 |
Standard lamp intensity analysis (1 sec, 1 rows)
| Serial | min | max | filter | fixable | 30-day history |
| #254 | 2568160 | 2570380 | 0 | yes |