Critical warnings from b-files' comments for days between 2023188 and 2023188 (1 sec, 4 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2023 187 05 44 36 Command not accepted while another in progress.11#25405:31:201-1
rl:2023 187 06 07 13 New value set for watermark LOOP.RUN.TIME: 700.000000 millisec.63#25405:31:281-1
rl:2023 187 06 07 13 New value set for watermark LOOP.RUN.TIME: 700.000000 millisec.63#25405:31:281-1
rl:2023 188 09 58 49 PMT counter failure11#25405:31:381-1

Instrument constant changes for the week before 2023188 (2 sec, 0 rows)

HG burned out? (AP results when HG is ON) (0 sec, 8 rows)

Non-critical warnings from b-files' comments for days between 2023188 and 2023188 (0 sec, 10 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
zs:ZS intensity too low ( 20/cy), skipping.299#25401:18:051-1
rl:2023 187 05 45 34 Bad command string. Command string buffer cleared.13#25405:31:231-1
rl:2023 187 05 45 34 Bad command string. Command string buffer cleared.13#25405:31:231-1
rl:2023 187 05 45 34 <13#25405:31:251-1
rl:2023 187 06 07 13 New value set for watermark LOOP.RUN.TIME: 700.000000 millisec.33#25405:31:281-1
rl:2023 187 06 07 13 New value set for watermark LOOP.RUN.TIME: 700.000000 millisec.33#25405:31:281-1
rl:2023 188 07 12 08 New value set for watermark LOOP.RUN.TIME: 719.000006 millisec.33#25405:31:361-1
rl:2023 188 07 12 08 New value set for watermark LOOP.RUN.TIME: 719.000006 millisec.33#25405:31:361-1
ds:DS intensity too low ( 133/cy), skipping.499#25410:52:481-1
sc:SC intensity too low ( 30/cy), skipping.13#25423:10:381-1

Measurements and tests summary.


_jday_loc_serial_skcdsdtedhghprssluvzs
188TORONTO#254calsc,Next schedule will be calsc (now calsc ZS= 90 ZS= 91.332),Next schedule will be calsc (now calsc ZS= 90 ZS= 91.902)71222928251717
(0 sec)

Number of days since last dome/window cleaning (0 sec, 1 rows)

Missing files (0 sec, 1 rows)

Seriallast datadays ago
2542023181007

Run/Stop results for last week. (0 sec, 1 rows)

SerialTotalFailed 0Failed darkFailed 1Failed 2Failed 3Failed 4Failed 5Failed double
2541606000000

AP results for SB Brewers. (0 sec, 1 rows)

SerialJdayt1 offt1 ont2 offt2 ont3 offt3 onHV offHV on+12V off+12V on+5V off+5V on-12V off-12V on+24V off+24V on+5V sps off+5V sps onSL A offSL A onSL V offSL V onWatts
254202318827.727.628.228.227.126.91389.41389.712.012.05.05.0-12.0-12.024.324.35.05.00.01.60.010.5016.8

Standard lamp intensity analysis (0 sec, 1 rows)

Serialminmaxfilterfixable30-day history
#254185760018586000