Database last updated on Sat, 12 Sep 2026 03:40:57 +0000 by localhost
Critical warnings from b-files' comments for days between 2019183 and 2019183 (0 sec, 4 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| rs: | Lamp not on... rs test terminated F1= 0 F4= 0 | 3 | 1 | B18319.246 | Delft | #246 | 08:09:20 | | 1 | -1 |
| hg: | Lamp not detected f1= 0 f4= 0 | 1 | 1 | B18319.246 | Delft | #246 | 15:13:34 | | 1 | -1 |
| rl: | 2019 183 02 02 10 PMT counter failure | 1 | 1 | B18319.246 | Delft | #246 | 23:46:59 | | 1 | -1 |
| rl: | 2019 183 08 27 28 Motor #11: appears stuck with a reference sensor activated. | 1 | 1 | B18319.246 | Delft | #246 | 23:47:02 | | 1 | -1 |
Instrument constant changes for the week before 2019183 (3 sec, 0 rows)
SL burned out? (AP results when SL is ON) (0 sec, 16 rows)
Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)
| Serial | YYYYJJJ | HG large | max correction |
| 246 | 2019183 | 2 | 5.2 |
Non-critical warnings from b-files' comments for days between 2019183 and 2019183 (0 sec, 9 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| hl: | HOME key pressed | 4 | 99 | B18319.246 | Delft | #246 | 07:43:18 | | 1 | -1 |
| hl: | Going to menu | 4 | 99 | B18319.246 | Delft | #246 | 07:43:19 | | 1 | -1 |
| sl: | DEL key pressed | 2 | 99 | B18319.246 | Delft | #246 | 07:47:30 | | 1 | -1 |
| rl: | 2019 182 23 53 43 Reset Warm -> Tepid from 2019 182 23 53 37 until 2019 182 23 53 37 . Counts: 1,0 | 1 | 3 | B18319.246 | Delft | #246 | 23:46:55 | | 1 | -1 |
| rl: | 2019 183 00 10 08 New value set for watermark LOOP.RUN.TIME: 700.000000 millisec. | 4 | 3 | B18319.246 | Delft | #246 | 23:46:56 | | 1 | -1 |
| rl: | 2019 183 00 10 08 New value set for watermark LOOP.RUN.TIME: 700.000000 millisec. | 4 | 3 | B18319.246 | Delft | #246 | 23:46:56 | | 1 | -1 |
| rl: | 2019 183 08 27 28 Motor #11: appears stuck with a reference sensor activated. | 1 | 3 | B18319.246 | Delft | #246 | 23:47:02 | | 1 | -1 |
| rl: | 2019 183 13 44 49 Reset Warm -> Warm from 2019 183 13 44 45 until 2019 183 13 44 46 . Counts: 0,1 | 1 | 3 | B18319.246 | Delft | #246 | 23:47:05 | | 1 | -1 |
| chk_cfg: | MOTOR.MAX.JERK[10]=(symbol not found) | 2 | 3 | B18319.246 | Delft | #246 | 23:47:13 | | 1 | -1 |
Measurements and tests summary.
| _jday | _loc | _serial | _skc | dt | ed | hg | hp | rs |
| 183 | DelftKZ | #246 | jawohl | 5 | 1 | 17 | 16 | 2 |
(0 sec)
Number of days since last dome/window cleaning (0 sec, 1 rows)
Missing files (0 sec, 1 rows)
| Serial | last data | days ago |
| 246 | 2019184 | 0-1 |
Questionable Run/Stop results. (0 sec, 1 rows)
| Serial | Jday | slit 0 | dark rs | dark counts | slit 1 | slit 2 | slit 3 | slit 4 | slit 5 | slits 3+5 |
| 246 | 2019183 | 0.998 | 1.026 | 476762 / 469878 (0.9M) | 0.954 | 0.988 | 0.991 | 0.994 | 0.996 | 0.992 |
Run/Stop results for last week. (0 sec, 1 rows)
| Serial | Total | Failed 0 | Failed dark | Failed 1 | Failed 2 | Failed 3 | Failed 4 | Failed 5 | Failed double |
| 246 | 40 | 14 | 8 | 17 | 17 | 17 | 14 | 12 | 16 |
Standard lamp intensity analysis (0 sec, 1 rows)
| Serial | min | max | filter | fixable | 30-day history |
| #246 | 1675000 | 2673310 | 0 | yes |