Database last updated on Sun, 01 Feb 2026 13:10:53 +0000 by localhost

Critical warnings from b-files' comments for days between 2024101 and 2024101 (0 sec, 1 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2024 101 05 00 07 Motor #3: appears stuck with a reference sensor activated.21#23906:28:321-1

Instrument constant changes for the week before 2024101 (1 sec, 0 rows)

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
2392024101110.2

Non-critical warnings from b-files' comments for days between 2024101 and 2024101 (0 sec, 4 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2024 101 05 00 07 Motor #3: appears stuck with a reference sensor activated.13#23906:28:321-1
rl:2024 101 05 00 08 Motor #11: appears stuck with a reference sensor activated.13#23906:28:341-1
ds:DS intensity too high ( 200680/cy) quitting.299#23916:11:251-1
ds:Suppressed oscillation of filter wheel.15#23916:11:451-1

Measurements and tests summary.


_jday_loc_serial_skcaodedsdtedfzhghprssluvzs
101CHURCHILL#239ecsc,Next schedule will be ecsc (now ecsc ZS= 93 ZS= 110.999),Next schedule will be ecsc (now ecsc ZS= 93 ZS= 93.93801)707222171716273140
(0 sec)

Number of days since last dome/window cleaning (0 sec, 3 rows)

serialdays
239more than a year
3more than a year

Missing files (0 sec, 1 rows)

Seriallast datadays ago
23920241020-1

Run/Stop results for last week. (0 sec, 1 rows)

SerialTotalFailed 0Failed darkFailed 1Failed 2Failed 3Failed 4Failed 5Failed double
2391603000000

AP results for SB Brewers. (0 sec, 1 rows)

SerialJdayt1 offt1 ont2 offt2 ont3 offt3 onHV offHV on+12V off+12V on+5V off+5V on-12V off-12V on+24V off+24V on+5V sps off+5V sps onSL A offSL A onSL V offSL V onWatts
239202410116.716.717.117.111.111.11351.31351.512.112.15.15.1-11.9-11.924.124.15.05.00.01.50.010.6515.8

Standard lamp intensity analysis (0 sec, 1 rows)

Serialminmaxfilterfixable30-day history
#239273142027361900yes