Database last updated on Mon, 07 Sep 2026 02:41:04 +0000 by localhost

Critical warnings from b-files' comments for days between 2022009 and 2022009 (0 sec, 1 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2022 008 06 39 11 Motor #11: appears stuck with a reference sensor activated.11B00922.239CHURC#23906:34:471-1

Instrument constant changes for the week before 2022009 (3 sec, 0 rows)

Non-critical warnings from b-files' comments for days between 2022009 and 2022009 (0 sec, 4 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2022 008 06 39 09 Reset Warm -> Tepid from 2022 008 06 39 04 until 2022 008 06 39 04 . Counts: 1,013B00922.239CHURC#23906:34:451-1
rl:2022 008 06 39 11 Motor #11: appears stuck with a reference sensor activated.13B00922.239CHURC#23906:34:471-1
rl:2022 009 07 16 32 New value set for watermark LOOP.RUN.TIME: 686.999989 millisec.13B00922.239CHURC#23906:34:501-1
rl:2022 009 07 16 32 New value set for watermark LOOP.RUN.TIME: 686.999989 millisec.13B00922.239CHURC#23906:34:501-1

Measurements and tests summary.


_jday_loc_serial_skcaodedsdtedfzhghprssluvzs
009CHURCHILL#239ecsc,Next schedule will be ecsc (now ecsc ZS= 93 ZS= 93.696)11111241716173611
(0 sec)

Number of days since last dome/window cleaning (0 sec, 3 rows)

serialdays
239more than a year
3more than a year

Missing files (0 sec, 1 rows)

Seriallast datadays ago
23920220100-1

Run/Stop results for last week. (0 sec, 1 rows)

SerialTotalFailed 0Failed darkFailed 1Failed 2Failed 3Failed 4Failed 5Failed double
239803000000

AP results for SB Brewers. (0 sec, 1 rows)

SerialJdayt1 offt1 ont2 offt2 ont3 offt3 onHV offHV on+12V off+12V on+5V off+5V on-12V off-12V on+24V off+24V on+5V sps off+5V sps onSL A offSL A onSL V offSL V onWatts
2392022009-0.4-0.40.50.5-14.5-14.61351.31351.312.012.05.05.1-12.0-12.024.724.35.05.00.01.50.010.9716.0

Standard lamp intensity analysis (0 sec, 1 rows)

Serialminmaxfilterfixable30-day history
#239217329021732900yes