Database last updated on Mon, 07 Sep 2026 11:41:00 +0000 by localhost

Critical warnings from b-files' comments for days between 2020319 and 2020319 (0 sec, 3 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
hp:Spectrometer alignment trouble.41B31920.239CHURC#23922:32:171-1
fr:11000 steps - QUIT - Micrometer #1 reference not found21B31920.239CHURC#23922:33:091-1
re:Previous RE didn't seem to help, so not doing reset again. Operator intervention required. Quitting to menu.11B31920.239CHURC#23922:48:591-1

Instrument constant changes for the week before 2020319 (2 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 3 rows)

Non-critical warnings from b-files' comments for days between 2020319 and 2020319 (0 sec, 1 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
re:Going to menu199B31920.239CHURC#23922:48:591-1

Measurements and tests summary.


_jday_loc_serial_skc
319CHURCHILL#239ecsc
(1 sec)

Number of days since last dome/window cleaning (0 sec, 3 rows)

serialdays
3more than a year
239343

Missing files (0 sec, 1 rows)

Seriallast datadays ago
23920203200-1

Run/Stop results for last week. (0 sec, 1 rows)

SerialTotalFailed 0Failed darkFailed 1Failed 2Failed 3Failed 4Failed 5Failed double
2391305000000

Ozone results for good DS obs (std<2.5DU, mu<4) within 3 minutes between 2 Brewers. (1 sec)

No ozone measurements to display.

Unsorted comments (0 sec, 579 rows)

SerialtypeTextCountYYYYJJJ
239env:DosBox12020319
239cubdsp:112020319
239ru:Giving SB time to complete motor movements.12020319
239en:NofUV12020319
239en:UA12020319
239en:UF12020319
239en:SC12020319
239en:ZS12020319
239en:ZP12020319
239en:ZUM12020319
239auto:Running:12020319
239td:Julian day = 31962020319

The END (0 sec, 2 rows)

If you have questions and/or comments about this page please contact Volodya Savastiouk