Database last updated on Wed, 09 Sep 2026 16:40:56 +0000 by localhost

Critical warnings from b-files' comments for days between 2019336 and 2019336 (0 sec, 1 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2019 335 06 09 28 Motor #11: appears stuck with a reference sensor activated.11B33619.239CHURC#23906:10:341-1

Instrument constant changes for the week before 2019336 (3 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 32 rows)

SerialYYYYJJJSL CurrentSL Voltage
2392019329-12.0124.25
2392019330-12.0124.22
2392019331-12.0124.25
2392019332-11.9924.25
2392019333-11.9924.22
2392019334-11.9924.22
2392019335-12.0124.25
2392019336-12.0124.22

Non-critical warnings from b-files' comments for days between 2019336 and 2019336 (0 sec, 7 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2019 335 06 09 26 Reset Warm -> Tepid from 2019 335 06 09 21 until 2019 335 06 09 21 . Counts: 1,013B33619.239CHURC#23906:10:331-1
rl:2019 335 06 09 28 Motor #11: appears stuck with a reference sensor activated.13B33619.239CHURC#23906:10:341-1
rl:2019 335 06 26 10 New value set for watermark LOOP.RUN.TIME: 699.000025 millisec.33B33619.239CHURC#23906:10:361-1
rl:2019 335 06 26 10 New value set for watermark LOOP.RUN.TIME: 699.000025 millisec.33B33619.239CHURC#23906:10:361-1
dp:Doing dispersion with internal HG.199B33619.239CHURC#23906:51:251-1
ds:DS intensity too low ( 11/cy), skipping.299B33619.239CHURC#23914:48:221-1
zs:ZS intensity too low ( 299/cy), skipping.299B33619.239CHURC#23914:48:341-1

Measurements and tests summary.


_jday_loc_serial_skcaodedsdtedfzhghprssluvzs
336CHURCHILL#239ecsc292921101312263213
(0 sec)

Number of days since last dome/window cleaning (0 sec, 3 rows)

serialdays
3more than a year
239025

Missing files (0 sec, 1 rows)

Seriallast datadays ago
23920193370-1

Run/Stop results for last week. (0 sec, 1 rows)

SerialTotalFailed 0Failed darkFailed 1Failed 2Failed 3Failed 4Failed 5Failed double
2391704000000

AP results for MB Brewer. (0 sec, 1 rows)

SerialJdayt1 offt1 ont2 offt2 ont3 offt3 onHV offHV on+15V off+15V on+5V off+5V on-15V off-15V on+24V off+24V on+5V sps off+5V sps on-8V sps off-8V sps onSL A offSL A onSL V offSL V onW
23920193368.30.08.00.04.61.31351.32.012.07.45.00.6-12.02.024.62.25.012.0-5.05.00.0-12.00.024.23-290.9

Standard lamp intensity analysis (0 sec, 1 rows)

Serialminmaxfilterfixable30-day history
#239212290021241700yes