Critical warnings from b-files' comments for days between 2019309 and 2019309 (0 sec, 1 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2019 308 06 07 57 Motor #11: appears stuck with a reference sensor activated.11B30919.239CHURC#23906:01:511-1

Instrument constant changes for the week before 2019309 (3 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 32 rows)

SerialYYYYJJJSL CurrentSL Voltage
2392019302-12.0124.22
2392019303-12.0124.22
2392019304-11.9924.25
2392019305-12.0124.25
2392019306-12.0124.25
2392019307-12.0124.25
2392019308-12.0124.22
2392019309-11.9924.22

Non-critical warnings from b-files' comments for days between 2019309 and 2019309 (0 sec, 5 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2019 308 06 07 56 Reset Warm -> Tepid from 2019 308 06 07 50 until 2019 308 06 07 50 . Counts: 1,013B30919.239CHURC#23906:01:501-1
rl:2019 308 06 07 57 Motor #11: appears stuck with a reference sensor activated.13B30919.239CHURC#23906:01:511-1
rl:2019 308 06 24 39 New value set for watermark LOOP.RUN.TIME: 699.000025 millisec.43B30919.239CHURC#23906:01:531-1
rl:2019 308 06 24 39 New value set for watermark LOOP.RUN.TIME: 699.000025 millisec.43B30919.239CHURC#23906:01:531-1
zs:ZS intensity too low ( 372/cy), skipping.499B30919.239CHURC#23922:12:281-1

Measurements and tests summary.


_jday_loc_serial_skcaodedsdtedfmfzhghprssluvzs
309CHURCHILL#239ecsc131321981110263339
(0 sec)

Number of days since last dome/window cleaning (0 sec, 3 rows)

serialdays
3more than a year
239029

Missing files (0 sec, 1 rows)

Seriallast datadays ago
23920193100-1

Run/Stop results for last week. (0 sec, 1 rows)

SerialTotalFailed 0Failed darkFailed 1Failed 2Failed 3Failed 4Failed 5Failed double
2391604000000

AP results for MB Brewer. (0 sec, 1 rows)

SerialJdayt1 offt1 ont2 offt2 ont3 offt3 onHV offHV on+15V off+15V on+5V off+5V on-15V off-15V on+24V off+24V on+5V sps off+5V sps on-8V sps off-8V sps onSL A offSL A onSL V offSL V onW
23920193097.90.07.70.01.21.31350.42.112.08.55.00.7-12.02.024.72.25.012.0-5.05.00.0-12.00.024.24-290.4

Standard lamp intensity analysis (0 sec, 1 rows)

Serialminmaxfilterfixable30-day history
#239211566021179300yes