Database last updated on Wed, 09 Sep 2026 20:11:06 +0000 by localhost

Critical warnings from b-files' comments for days between 2019308 and 2019308 (0 sec, 1 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2019 307 06 05 17 Motor #11: appears stuck with a reference sensor activated.11B30819.239CHURC#23906:04:031-1

Instrument constant changes for the week before 2019308 (3 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 32 rows)

SerialYYYYJJJSL CurrentSL Voltage
2392019301-11.9924.25
2392019302-12.0124.22
2392019303-12.0124.22
2392019304-11.9924.25
2392019305-12.0124.25
2392019306-12.0124.25
2392019307-12.0124.25
2392019308-12.0124.22

Non-critical warnings from b-files' comments for days between 2019308 and 2019308 (0 sec, 6 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2019 307 06 05 16 Reset Warm -> Tepid from 2019 307 06 05 10 until 2019 307 06 05 11 . Counts: 1,013B30819.239CHURC#23906:04:021-1
rl:2019 307 06 05 17 Motor #11: appears stuck with a reference sensor activated.13B30819.239CHURC#23906:04:031-1
rl:2019 307 06 21 58 New value set for watermark LOOP.RUN.TIME: 700.000000 millisec.33B30819.239CHURC#23906:04:051-1
rl:2019 307 06 21 58 New value set for watermark LOOP.RUN.TIME: 700.000000 millisec.33B30819.239CHURC#23906:04:051-1
rl:2019 308 13 57 52 New value set for watermark LOOP.RUN.TIME: 716.999960 millisec.13B30819.239CHURC#23906:04:091-1
rl:2019 308 13 57 52 New value set for watermark LOOP.RUN.TIME: 716.999960 millisec.13B30819.239CHURC#23906:04:091-1

Measurements and tests summary.


_jday_loc_serial_skcaodedsdtedfzhghprssluvzs
308CHURCHILL#239ecsc1212216109273636
(0 sec)

Number of days since last dome/window cleaning (0 sec, 3 rows)

serialdays
3more than a year
239028

Missing files (0 sec, 1 rows)

Seriallast datadays ago
23920193090-1

Run/Stop results for last week. (0 sec, 1 rows)

SerialTotalFailed 0Failed darkFailed 1Failed 2Failed 3Failed 4Failed 5Failed double
2391605000000

AP results for MB Brewer. (0 sec, 1 rows)

SerialJdayt1 offt1 ont2 offt2 ont3 offt3 onHV offHV on+15V off+15V on+5V off+5V on-15V off-15V on+24V off+24V on+5V sps off+5V sps on-8V sps off-8V sps onSL A offSL A onSL V offSL V onW
23920193087.90.07.90.03.21.31350.72.012.08.45.00.7-12.02.024.62.25.012.0-5.05.00.0-12.00.024.22-290.9

Standard lamp intensity analysis (0 sec, 1 rows)

Serialminmaxfilterfixable30-day history
#239210723021097900yes