Critical warnings from b-files' comments for days between 2026202 and 2026202 (1 sec, 3 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
hp:Spectrometer alignment trouble.31B20226.235RESOL#23509:37:341-1
re:Previous RE didn't seem to help, so not doing reset again. Operator intervention required. Quitting to menu.11B20226.235RESOL#23509:45:501-1
bo:Command file bo not found --- continuing11B20226.235RESOL#23514:25:061-1

Instrument constant changes for the week before 2026202 (2 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 2 rows)

SerialYYYYJJJSL CurrentSL Voltage
23520262011.525.79
23520262021.526.05

Non-critical warnings from b-files' comments for days between 2026202 and 2026202 (0 sec, 5 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
skc:Going to menu399B20226.235RESOL#23509:45:561-1
hp:HOME key pressed399B20226.235RESOL#23513:26:091-1
hh:CALCULATED M1<0 SET TO 0299B20226.235RESOL#23513:36:2501-1
hh:CALCULATED M1<0 SET TO 023B20226.235RESOL#23513:36:251-1
bo:Command file bo not found --- continuing199B20226.235RESOL#23514:25:06continuing1-1

Measurements and tests summary.


_jday_loc_serial_skcdsdtfzhghprsslzs
202RESOLUTE#2350 0 2967.28 0.0 0 18,Attempting recovery RESET,ecnouv,Going to menu,Must do HG now (C1414111119
(0 sec)

Number of days since last dome/window cleaning (0 sec, 3 rows)

serialdays
235more than a year
3more than a year

Missing files (0 sec, 1 rows)

Seriallast datadays ago
23520262030-1

Questionable Run/Stop results. (0 sec, 1 rows)

SerialJdayslit 0dark rsdark countsslit 1slit 2slit 3slit 4slit 5slits 3+5
23520262020.9911.0001 / 0 (0.3M)0.9830.9880.9900.9880.9960.991

Run/Stop results for last week. (0 sec, 1 rows)

SerialTotalFailed 0Failed darkFailed 1Failed 2Failed 3Failed 4Failed 5Failed double
235820211211

AP results for SB Brewers. (0 sec, 1 rows)

SerialJdayt1 offt1 ont2 offt2 ont3 offt3 onHV offHV on+12V off+12V on+5V off+5V on-12V off-12V on+24V off+24V on+5V sps off+5V sps onSL A offSL A onSL V offSL V onWatts
235202620218.017.917.717.615.815.81453.31452.312.012.05.05.0-11.9-11.924.024.05.05.00.00.83.410.234.6

Standard lamp intensity analysis (0 sec, 1 rows)

Serialminmaxfilterfixable30-day history
#2354653844653840