Database last updated on Tue, 03 Feb 2026 02:40:51 +0000 by localhost
Instrument constant changes for the week before 2023170 (2 sec, 0 rows)
HG burned out? (AP results when HG is ON) (0 sec, 7 rows)
Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)
| Serial | YYYYJJJ | HG large | max correction |
| 235 | 2023170 | 2 | 2.8 |
Non-critical warnings from b-files' comments for days between 2023170 and 2023170 (0 sec, 2 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| si: | New SI, HC = 5, NC = 570 Old = 9 , 575 | 1 | 99 | B17023.235 | RESOL | #235 | 03:34:58 | | 1 | -1 |
| zs: | ZS measurement terminated | 1 | 5 | B17023.235 | RESOL | #235 | 13:25:29 | | 1 | -1 |
Measurements and tests summary.
| _jday | _loc | _serial | _skc | dt | hg | hp | rs | sl | zs |
| 170 | RESOLUTE | #235 | calds | 5 | 22 | 17 | 6 | 14 | 58 |
(0 sec)
Number of days since last dome/window cleaning (0 sec, 3 rows)
| serial | days |
| 235 | more than a year |
| 3 | more than a year |
Missing files (0 sec, 1 rows)
| Serial | last data | days ago |
| 235 | 2023171 | 0-1 |
Questionable Run/Stop results. (0 sec, 1 rows)
| Serial | Jday | slit 0 | dark rs | dark counts | slit 1 | slit 2 | slit 3 | slit 4 | slit 5 | slits 3+5 |
| 235 | 2023170 | 0.999 | | 0 / 1 (0.2M) | 0.984 | 0.981 | 0.990 | 0.983 | 0.989 | 0.985 |
Run/Stop results for last week. (0 sec, 1 rows)
| Serial | Total | Failed 0 | Failed dark | Failed 1 | Failed 2 | Failed 3 | Failed 4 | Failed 5 | Failed double |
| 235 | 13 | 4 | 7 | 4 | 5 | 5 | 4 | 3 | 3 |
AP results for SB Brewers. (0 sec, 1 rows)
| Serial | Jday | t1 off | t1 on | t2 off | t2 on | t3 off | t3 on | HV off | HV on | +12V off | +12V on | +5V off | +5V on | -12V off | -12V on | +24V off | +24V on | +5V sps off | +5V sps on | SL A off | SL A on | SL V off | SL V on | Watts |
| 235 | 2023170 | 17.9 | 17.9 | 18.5 | 18.5 | 17.6 | 17.6 | 1427.7 | 1427.7 | 12.0 | 12.0 | 5.0 | 5.0 | -11.9 | -11.9 | 24.6 | 24.1 | 5.0 | 5.0 | 0.0 | 1.5 | 0.0 | 9.65 | 14.6 |
Standard lamp intensity analysis (0 sec, 1 rows)
| Serial | min | max | filter | fixable | 30-day history |
| #235 | 355214 | 1293380 | 0 | |