Critical warnings from b-files' comments for days between 2020046 and 2020046 (0 sec, 1 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2020 046 10 42 32 Motor #3: appears stuck with a reference sensor activated.11B04620.235RESOL#23506:39:301-1

Instrument constant changes for the week before 2020046 (3 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 19 rows)

SerialYYYYJJJSL CurrentSL Voltage
2352020039-11.8824.25
2352020040-11.8824.25
2352020041-11.8324.31
2352020042-11.8224.37
2352020043-11.8624.28
2352020044-11.8624.28
2352020045-11.8624.25
2352020046-11.8324.28

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
2352020046115.3

Non-critical warnings from b-files' comments for days between 2020046 and 2020046 (0 sec, 4 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2020 046 10 42 31 Reset Tepid -> Tepid from 2020 046 10 42 09 until 2020 046 10 42 26 . Counts: 1,013B04620.235RESOL#23506:39:281-1
rl:2020 046 10 42 32 Motor #3: appears stuck with a reference sensor activated.13B04620.235RESOL#23506:39:301-1
rl:2020 046 20 10 12 New value set for watermark LOOP.RUN.TIME: 671.999979 millisec.23B04620.235RESOL#23506:39:321-1
rl:2020 046 20 10 12 New value set for watermark LOOP.RUN.TIME: 671.999979 millisec.23B04620.235RESOL#23506:39:321-1

Measurements and tests summary.


_jday_loc_serial_skcaodedsdtedfzhghprssluvzs
046RESOLUTE#235ecsc2231201416363111
(0 sec)

Number of days since last dome/window cleaning (0 sec, 3 rows)

serialdays
3more than a year
235070

Missing files (0 sec, 1 rows)

Seriallast datadays ago
23520200470-1

Run/Stop results for last week. (0 sec, 1 rows)

SerialTotalFailed 0Failed darkFailed 1Failed 2Failed 3Failed 4Failed 5Failed double
2351807000000

AP results for MB Brewer. (0 sec, 1 rows)

SerialJdayt1 offt1 ont2 offt2 ont3 offt3 onHV offHV on+15V off+15V on+5V off+5V on-15V off-15V on+24V off+24V on+5V sps off+5V sps on-8V sps off-8V sps onSL A offSL A onSL V offSL V onW
23520200462.10.02.30.0-12.81.21427.81.912.08.35.00.5-11.82.024.82.25.012.0-4.85.00.0-11.80.024.28-287.2

Standard lamp intensity analysis (0 sec, 1 rows)

Serialminmaxfilterfixable30-day history
#235204028020470700yes