Critical warnings from b-files' comments for days between 2019115 and 2019115 (0 sec, 2 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2019 114 05 45 39 New value set for watermark LOOP.RUN.TIME: 688.000011 millisec.63B11519.235TORON#23505:18:581-1
rl:2019 114 05 45 39 New value set for watermark LOOP.RUN.TIME: 688.000011 millisec.63B11519.235TORON#23505:18:581-1

Instrument constant changes for the week before 2019115 (3 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 28 rows)

SerialYYYYJJJSL CurrentSL Voltage
2352019108-11.8924.22
2352019109-11.924.22
2352019110-11.8824.22
2352019111-11.8924.22
2352019112-11.8924.22
2352019113-11.8924.2
2352019114-11.8824.22
2352019115-11.8824.22

Non-critical warnings from b-files' comments for days between 2019115 and 2019115 (0 sec, 5 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2019 114 05 27 58 Reset Warm -> Tepid from 2019 114 05 27 52 until 2019 114 05 27 53 . Counts: 1,013B11519.235TORON#23505:18:551-1
rl:2019 115 13 34 41 New value set for watermark LOOP.RUN.TIME: 721.000004 millisec.13B11519.235TORON#23505:19:091-1
rl:2019 115 13 34 41 New value set for watermark LOOP.RUN.TIME: 721.000004 millisec.13B11519.235TORON#23505:19:091-1
ds:Suppressed oscillation of filter wheel.25B11519.235TORON#23512:09:531-1
ds:DS intensity too high ( 202342/cy) quitting.299B11519.235TORON#23512:46:221-1

Measurements and tests summary.


_jday_loc_serial_skcaodedsdtedfmfzhghprssluvzs
115TORONTO#235ecauto575911621312145027
(0 sec)

Number of days since last dome/window cleaning (0 sec, 3 rows)

serialdays
3more than a year
235018

Missing files (0 sec, 1 rows)

Seriallast datadays ago
23520191160-1

Run/Stop results for last week. (0 sec, 1 rows)

SerialTotalFailed 0Failed darkFailed 1Failed 2Failed 3Failed 4Failed 5Failed double
235803000000

AP results for MB Brewer. (0 sec, 1 rows)

SerialJdayt1 offt1 ont2 offt2 ont3 offt3 onHV offHV on+15V off+15V on+5V off+5V on-15V off-15V on+24V off+24V on+5V sps off+5V sps on-8V sps off-8V sps onSL A offSL A onSL V offSL V onW
235201911513.70.013.30.012.21.31427.22.112.03.85.00.4-11.92.024.62.25.012.0-4.95.00.0-11.90.024.22-287.7

Standard lamp intensity analysis (0 sec, 1 rows)

Serialminmaxfilterfixable30-day history
#235200728020072800yes