Database last updated on Sun, 13 Sep 2026 09:10:59 +0000 by localhost

Critical warnings from b-files' comments for days between 2018360 and 2018360 (0 sec, 2 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2018 359 05 48 00 New value set for watermark LOOP.RUN.TIME: 699.000025 millisec.83B36018.235TORON#23505:29:391-1
rl:2018 359 05 48 00 New value set for watermark LOOP.RUN.TIME: 699.000025 millisec.83B36018.235TORON#23505:29:391-1

Instrument constant changes for the week before 2018360 (1 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 8 rows)

HG burned out? (AP results when HG is ON) (0 sec, 11 rows)

SerialYYYYJJJHG CurrentHG Voltage
23520183530.1714.94
23520183540.1714.95
23520183550.1714.97
23520183560.1714.92
23520183570.1714.94
23520183580.1714.94
23520183590.3213.6
23520183590.1714.94
23520183590.1714.95
23520183600.3213.6
23520183600.1714.94

Non-critical warnings from b-files' comments for days between 2018360 and 2018360 (0 sec, 4 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2018 359 05 30 21 Reset Warm -> Tepid from 2018 359 05 30 15 until 2018 359 05 30 16 . Counts: 1,013B36018.235TORON#23505:29:371-1
rl:2018 360 13 39 59 New value set for watermark LOOP.RUN.TIME: 718.000031 millisec.33B36018.235TORON#23505:29:511-1
rl:2018 360 13 39 59 New value set for watermark LOOP.RUN.TIME: 718.000031 millisec.33B36018.235TORON#23505:29:511-1
zs:ZS intensity too low ( 341/cy), skipping.199B36018.235TORON#23513:09:591-1

Measurements and tests summary.


_jday_loc_serial_skcaodedsdtedfmfzhghprssluvzs
360TORONTO#235ecsc292921283109253329
(0 sec)

New operators? (1 sec, 18 rows)

Number of days since last dome/window cleaning (0 sec, 3 rows)

serialdays
3more than a year
235019

Missing files (0 sec, 1 rows)

Seriallast datadays ago
23520183610-1

Run/Stop results for last week. (0 sec, 1 rows)

SerialTotalFailed 0Failed darkFailed 1Failed 2Failed 3Failed 4Failed 5Failed double
2351607000000

AP results for SB Brewers. (0 sec, 1 rows)

SerialJdayt1 offt1 ont2 offt2 ont3 offt3 onHV offHV on+12V off+12V on+5V off+5V on-12V off-12V on+24V off+24V on+5V sps off+5V sps onSL A offSL A onSL V offSL V onWatts
23520183608.48.48.38.37.77.71427.21425.912.012.05.05.0-11.9-11.924.724.25.05.00.01.50.010.4915.3

Standard lamp intensity analysis (0 sec, 1 rows)

Serialminmaxfilterfixable30-day history
#235200862020139300yes