Database last updated on Mon, 26 Jan 2026 19:10:52 +0000 by localhost

Critical warnings from b-files' comments for days between 2025124 and 2025124 (0 sec, 1 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2025 123 05 34 01 Motor #2: appears stuck with a reference sensor activated.21B12425.224TORON#22405:22:051-1

Instrument constant changes for the week before 2025124 (2 sec, 0 rows)

Non-critical warnings from b-files' comments for days between 2025124 and 2025124 (1 sec, 4 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2025 123 05 31 32 Reset Warm -> Tepid from 2025 123 05 31 26 until 2025 123 05 31 27 . Counts: 1,013B12425.224TORON#22405:22:031-1
rl:2025 124 14 26 02 New value set for watermark SUPPLY.VOLTAGE.HIGH43B12425.224TORON#22405:22:111-1
ds:DS intensity too high ( 208565/cy) quitting.499B12425.224TORON#22413:08:571-1
zs:ZS is not done when solar zenith angle less than 30 degrees. Skipping.110B12425.224TORON#22416:20:221-1

Measurements and tests summary.


_jday_loc_serial_skcdsdtedfzhgrssluvzs
124TORONTO#224ecsc,Next schedule will be ecsc (now ecsc ZS= 93 ZS= 93.322)6322721263052
(1 sec)

Number of days since last dome/window cleaning (1 sec, 2 rows)

serialdays
224more than a year

Missing files (0 sec, 1 rows)

Seriallast datadays ago
22420251250-1

Run/Stop results for last week. (0 sec, 1 rows)

SerialTotalFailed 0Failed darkFailed 1Failed 2Failed 3Failed 4Failed 5Failed double
2241601000000

Standard lamp intensity analysis (0 sec, 1 rows)

Serialminmaxfilterfixable30-day history
#224266986026703600yes