Critical warnings from b-files' comments for days between 2025084 and 2025084 (0 sec, 1 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| rl: | 2025 084 05 42 02 Motor #2: appears stuck with a reference sensor activated. | 3 | 1 | B08425.224 | TORON | #224 | 05:29:51 | | 1 | -1 |
Instrument constant changes for the week before 2025084 (3 sec, 0 rows)
SL burned out? (AP results when SL is ON) (0 sec, 2 rows)
Non-critical warnings from b-files' comments for days between 2025084 and 2025084 (0 sec, 4 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| rl: | 2025 084 05 39 20 Reset Warm -> Tepid from 2025 084 05 39 14 until 2025 084 05 39 14 . Counts: 1,0 | 1 | 3 | B08425.224 | TORON | #224 | 05:29:48 | | 1 | -1 |
| re: | Azimuth discrepancy = -195 | 1 | 99 | B08425.224 | TORON | #224 | 05:42:08 | -195 | -30 | 30 |
| ds: | Suppressed oscillation of filter wheel. | 1 | 5 | B08425.224 | TORON | #224 | 15:06:11 | | 1 | -1 |
| ds: | DS intensity too high ( 306404/cy) quitting. | 1 | 99 | B08425.224 | TORON | #224 | 18:15:22 | | 1 | -1 |
Measurements and tests summary.
| _jday | _loc | _serial | _skc | ds | dt | ed | fz | hg | hp | rs | sl | uv | zs |
| 084 | TORONTO | #224 | ecsc,Next schedule will be ecsc (now ecsc ZS= 93 ZS= 93.04) | 28 | 2 | 2 | 7 | 18 | 19 | 2 | 6 | 25 | 65 |
(0 sec)
Number of days since last dome/window cleaning (0 sec, 2 rows)
| serial | days |
| 224 | more than a year |
Missing files (0 sec, 1 rows)
| Serial | last data | days ago |
| 224 | 2025085 | 0-1 |
Run/Stop results for last week. (0 sec, 1 rows)
| Serial | Total | Failed 0 | Failed dark | Failed 1 | Failed 2 | Failed 3 | Failed 4 | Failed 5 | Failed double |
| 224 | 15 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 |
AP results for SB Brewers. (0 sec, 1 rows)
| Serial | Jday | t1 off | t1 on | t2 off | t2 on | t3 off | t3 on | HV off | HV on | +12V off | +12V on | +5V off | +5V on | -12V off | -12V on | +24V off | +24V on | +5V sps off | +5V sps on | SL A off | SL A on | SL V off | SL V on | Watts |
| 224 | 2025084 | 13.4 | 13.3 | 18.4 | 18.2 | 11.1 | 11.1 | 1275.7 | 1274.6 | 12.0 | 12.0 | 5.2 | 5.2 | -12.4 | -12.4 | 24.1 | 24.1 | 5.0 | 5.0 | 0.0 | 1.5 | 0.0 | 10.47 | 15.8 |
Standard lamp intensity analysis (0 sec, 1 rows)
| Serial | min | max | filter | fixable | 30-day history |
| #224 | 2670600 | 2671970 | 0 | yes |