Database last updated on Sun, 01 Feb 2026 12:40:51 +0000 by localhost

Critical warnings from b-files' comments for days between 2024111 and 2024111 (0 sec, 3 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2024 110 05 31 37 Motor #2: appears stuck with a reference sensor activated.21#22405:31:281-1
rl:2024 111 17 39 39 New value set for watermark SUPPLY.VOLTAGE.LOW[93#22405:31:331-1
sc:SC intensity too low ( 127/cy), skipping.93#22412:22:371-1

Instrument constant changes for the week before 2024111 (1 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 2 rows)

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
2242024111113.5

Non-critical warnings from b-files' comments for days between 2024111 and 2024111 (0 sec, 1 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
ds:DS intensity too high ( 402217/cy) quitting.299#22418:26:361-1

Measurements and tests summary.


_jday_loc_serial_skcdsdtedfmfzhghprssluvzs
111TORONTO#224calsc,Next schedule will be calsc (now calsc ZS= 90 ZS= 91.038),Next schedule will be calsc (now calsc ZS= 90 ZS= 91.675)87222032323261618
(0 sec)

Number of days since last dome/window cleaning (0 sec, 2 rows)

serialdays
224more than a year

Run/Stop results for last week. (0 sec, 1 rows)

SerialTotalFailed 0Failed darkFailed 1Failed 2Failed 3Failed 4Failed 5Failed double
224800000000

AP results for SB Brewers. (0 sec, 1 rows)

SerialJdayt1 offt1 ont2 offt2 ont3 offt3 onHV offHV on+12V off+12V on+5V off+5V on-12V off-12V on+24V off+24V on+5V sps off+5V sps onSL A offSL A onSL V offSL V onWatts
224202411116.015.920.120.012.512.51316.01315.112.012.05.15.1-12.1-12.124.124.15.05.00.01.50.010.7916.3

Standard lamp intensity analysis (0 sec, 1 rows)

Serialminmaxfilterfixable30-day history
#224392422039246800yes