Database last updated on Sun, 01 Feb 2026 23:10:52 +0000 by localhost
Critical warnings from b-files' comments for days between 2023352 and 2023352 (1 sec, 1 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| rl: | 2023 352 05 30 21 Motor #2: appears stuck with a reference sensor activated. | 4 | 1 | | | #224 | 05:21:33 | | 1 | -1 |
Instrument constant changes for the week before 2023352 (1 sec, 0 rows)
HG burned out? (AP results when HG is ON) (0 sec, 9 rows)
Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)
| Serial | YYYYJJJ | HG large | max correction |
| 224 | 2023352 | 4 | 3.3 |
Non-critical warnings from b-files' comments for days between 2023352 and 2023352 (1 sec, 5 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| rl: | 2023 352 05 31 32 Bad command string. Command string buffer cleared. | 2 | 3 | | | #224 | 05:21:35 | | 1 | -1 |
| rl: | 2023 352 05 31 32 Bad command string. Command string buffer cleared. | 2 | 3 | | | #224 | 05:21:35 | | 1 | -1 |
| rl: | 2023 352 05 31 33 < | 2 | 3 | | | #224 | 05:21:37 | | 1 | -1 |
| rl: | 2023 352 02 28 22 New value set for watermark SUPPLY.VOLTAGE.HIGH[ | 2 | 3 | | | #224 | 05:21:41 | | 1 | -1 |
| dp: | Doing dispersion with internal HG. | 2 | 99 | | | #224 | 10:42:02 | | 1 | -1 |
Measurements and tests summary.
| _jday | _loc | _serial | _skc | ds | dt | ed | fz | hg | hp | rs | sl | uv | zs |
| 352 | TORONTO,TORONTO | #224 | calsc,Next schedule will be calsc (now calsc ZS= 90 ZS= 90.903),Next schedule will be calsc (now calsc ZS= 90 ZS= 91.455),Waiting until ZA = | 13 | 4 | 2 | 6 | 32 | 16 | 4 | 6 | 19 | 48 |
(1 sec)
Number of days since last dome/window cleaning (0 sec, 2 rows)
| serial | days |
| 224 | more than a year |
Run/Stop results for last week. (0 sec, 1 rows)
| Serial | Total | Failed 0 | Failed dark | Failed 1 | Failed 2 | Failed 3 | Failed 4 | Failed 5 | Failed double |
| 224 | 18 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 |
AP results for SB Brewers. (0 sec, 1 rows)
| Serial | Jday | t1 off | t1 on | t2 off | t2 on | t3 off | t3 on | HV off | HV on | +12V off | +12V on | +5V off | +5V on | -12V off | -12V on | +24V off | +24V on | +5V sps off | +5V sps on | SL A off | SL A on | SL V off | SL V on | Watts |
| 224 | 2023352 | 17.3 | 17.3 | 18.6 | 18.6 | 16.7 | 16.7 | 1164.6 | 1163.5 | 12.0 | 12.0 | 5.2 | 5.1 | -12.2 | -12.2 | 24.1 | 24.1 | 5.0 | 5.0 | 0.0 | 1.5 | 0.0 | 10.70 | 16.2 |
Standard lamp intensity analysis (0 sec, 1 rows)
| Serial | min | max | filter | fixable | 30-day history |
| #224 | 2369660 | 2371910 | 0 | yes |