Critical warnings from b-files' comments for days between 2023341 and 2023341 (0 sec, 2 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| rl: | 2023 340 05 23 53 Motor #2: appears stuck with a reference sensor activated. | 1 | 1 | B34123.224 | TORON | #224 | 05:12:44 | | 1 | -1 |
| pd: | Brewer failed to respond 5 times | 1 | 1 | B34123.224 | TORON | #224 | 05:26:20 | | 1 | -1 |
Instrument constant changes for the week before 2023341 (2 sec, 0 rows)
HG burned out? (AP results when HG is ON) (0 sec, 16 rows)
Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)
| Serial | YYYYJJJ | HG large | max correction |
| 224 | 2023341 | 3 | 1029.0 |
HP with step corrections larger than 20 steps (0 sec, 1 rows)
| Serial | YYYYJJJ | HP large | max correction |
| 224 | 2023341 | 1 | 21.2 |
Non-critical warnings from b-files' comments for days between 2023341 and 2023341 (0 sec, 5 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| rl: | 2023 340 14 35 14 Bad command string. Command string buffer cleared. | 1 | 3 | B34123.224 | TORON | #224 | 05:12:47 | | 1 | -1 |
| rl: | 2023 340 14 35 14 Bad command string. Command string buffer cleared. | 1 | 3 | B34123.224 | TORON | #224 | 05:12:47 | | 1 | -1 |
| rl: | 2023 340 14 35 14 < | 1 | 3 | B34123.224 | TORON | #224 | 05:12:49 | | 1 | -1 |
| sc: | SC intensity too low ( 119/cy), skipping. | 4 | 3 | B34123.224 | TORON | #224 | 14:39:59 | | 1 | -1 |
| zs: | ZS intensity too low ( 3/cy), skipping. | 3 | 99 | B34123.224 | TORON | #224 | 22:10:05 | | 1 | -1 |
Measurements and tests summary.
| _jday | _loc | _serial | _skc | ds | ed | hg | hp | sl | uv | zs |
| 341 | TORONTO | #224 | calsc,Next schedule will be calsc (now calsc ZS= 90 ZS= 90.838),Next schedule will be calsc (now calsc ZS= 90 ZS= 91.386) | 34 | 2 | 11 | 9 | 4 | 8 | 26 |
(0 sec)
Number of days since last dome/window cleaning (0 sec, 2 rows)
| serial | days |
| 224 | more than a year |
Run/Stop results for last week. (0 sec, 1 rows)
| Serial | Total | Failed 0 | Failed dark | Failed 1 | Failed 2 | Failed 3 | Failed 4 | Failed 5 | Failed double |
| 224 | 15 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 |
AP results for SB Brewers. (0 sec, 1 rows)
| Serial | Jday | t1 off | t1 on | t2 off | t2 on | t3 off | t3 on | HV off | HV on | +12V off | +12V on | +5V off | +5V on | -12V off | -12V on | +24V off | +24V on | +5V sps off | +5V sps on | SL A off | SL A on | SL V off | SL V on | Watts |
| 224 | 2023341 | 16.2 | 16.1 | 18.2 | 18.1 | 14.9 | 15.0 | 1165.0 | 1163.1 | 11.9 | 12.0 | 5.2 | 5.1 | -12.3 | -12.3 | 24.1 | 24.1 | 5.0 | 5.0 | 0.0 | 1.5 | 0.0 | 10.68 | 16.1 |
Suspicious HG test results (correlation<0.98 or counts<5000). (0 sec, 1 rows)
| Serial | Jday | count | Peak | min.corr. | max.corr. | 30-day history |
| #224 | 341 | 2 | 143 | 0 | 0 |