Database last updated on Tue, 01 Sep 2026 22:11:04 +0000 by localhost

Critical warnings from b-files' comments for days between 2023333 and 2023333 (0 sec, 1 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2023 333 05 24 46 Motor #2: appears stuck with a reference sensor activated.21#22405:21:151-1

Instrument constant changes for the week before 2023333 (3 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 3 rows)

HG burned out? (AP results when HG is ON) (0 sec, 7 rows)

Non-critical warnings from b-files' comments for days between 2023333 and 2023333 (0 sec, 6 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2023 333 05 26 26 Bad command string. Command string buffer cleared.13#22405:21:171-1
rl:2023 333 05 26 26 Bad command string. Command string buffer cleared.13#22405:21:171-1
rl:2023 333 05 26 26 <13#22405:21:191-1
rl:2023 333 07 07 55 New value set for watermark SUPPLY.VOLTAGE.HIGH[13#22405:21:241-1
dp:Doing dispersion with internal HG.199#22411:12:581-1
sc:SC intensity too low ( 234/cy), skipping.33#22417:12:481-1

Measurements and tests summary.


_jday_loc_serial_skcdsdtedfmhghprssluvzs
333TORONTO#224calsc,Next schedule will be calsc (now calsc ZS= 90 ZS= 90.63201),Next schedule will be calsc (now calsc ZS= 90 ZS= 91.2),Waiting until ZA =3422171616261334
(0 sec)

Number of days since last dome/window cleaning (0 sec, 2 rows)

serialdays
224more than a year

Run/Stop results for last week. (0 sec, 1 rows)

SerialTotalFailed 0Failed darkFailed 1Failed 2Failed 3Failed 4Failed 5Failed double
2242001000000

AP results for SB Brewers. (0 sec, 1 rows)

SerialJdayt1 offt1 ont2 offt2 ont3 offt3 onHV offHV on+12V off+12V on+5V off+5V on-12V off-12V on+24V off+24V on+5V sps off+5V sps onSL A offSL A onSL V offSL V onWatts
224202333314.214.217.617.611.711.71165.01163.912.012.05.15.1-12.1-12.124.124.15.05.00.01.50.010.6716.1

Standard lamp intensity analysis (0 sec, 1 rows)

Serialminmaxfilterfixable30-day history
#224235467023603700yes