Database last updated on Wed, 02 Sep 2026 01:41:01 +0000 by localhost
Critical warnings from b-files' comments for days between 2023251 and 2023251 (0 sec, 3 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| rl: | 2023 250 05 32 47 Motor #2: appears stuck with a reference sensor activated. | 2 | 1 | | | #224 | 05:24:09 | | 1 | -1 |
| rl: | 2023 251 07 35 52 New value set for watermark SUPPLY.VOLTAGE.LOW[ | 6 | 3 | | | #224 | 05:24:16 | | 1 | -1 |
| sc: | SC intensity too low ( 66/cy), skipping. | 5 | 3 | | | #224 | 12:15:49 | | 1 | -1 |
Instrument constant changes for the week before 2023251 (3 sec, 0 rows)
HG burned out? (AP results when HG is ON) (0 sec, 19 rows)
HP with step corrections larger than 20 steps (0 sec, 1 rows)
| Serial | YYYYJJJ | HP large | max correction |
| 224 | 2023251 | 1 | 27.3 |
Non-critical warnings from b-files' comments for days between 2023251 and 2023251 (0 sec, 2 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| dp: | Doing dispersion with internal HG. | 1 | 99 | | | #224 | 09:48:41 | | 1 | -1 |
| ds: | DS intensity too low ( 5/cy), skipping. | 2 | 99 | | | #224 | 11:09:52 | | 1 | -1 |
Measurements and tests summary.
| _jday | _loc | _serial | _skc | ds | dt | ed | fz | hg | hp | rs | sl | uv | zs |
| 251 | TORONTO | #224 | calsc,Next schedule will be calsc (now calsc ZS= 90 ZS= 90.572),Waiting until ZA = | 60 | 2 | 2 | 2 | 21 | 22 | 2 | 6 | 15 | 28 |
(0 sec)
Number of days since last dome/window cleaning (0 sec, 2 rows)
| serial | days |
| 224 | more than a year |
Missing files (0 sec, 1 rows)
| Serial | last data | days ago |
| 224 | 2023244 | 007 |
Questionable Run/Stop results. (0 sec, 1 rows)
| Serial | Jday | slit 0 | dark rs | dark counts | slit 1 | slit 2 | slit 3 | slit 4 | slit 5 | slits 3+5 |
| 224 | 2023251 | 0.978 | 1.171 | 41 / 35 (1.5M) | 0.970 | 0.946 | 0.948 | 0.955 | 0.992 | 0.977 |
Run/Stop results for last week. (0 sec, 1 rows)
| Serial | Total | Failed 0 | Failed dark | Failed 1 | Failed 2 | Failed 3 | Failed 4 | Failed 5 | Failed double |
| 224 | 15 | 8 | 0 | 9 | 9 | 9 | 8 | 8 | 10 |
AP results for SB Brewers. (0 sec, 1 rows)
| Serial | Jday | t1 off | t1 on | t2 off | t2 on | t3 off | t3 on | HV off | HV on | +12V off | +12V on | +5V off | +5V on | -12V off | -12V on | +24V off | +24V on | +5V sps off | +5V sps on | SL A off | SL A on | SL V off | SL V on | Watts |
| 224 | 2023251 | 19.7 | 19.7 | 20.6 | 20.6 | 19.5 | 19.5 | 1168.7 | 1168.9 | 12.0 | 12.0 | 5.2 | 5.2 | -11.8 | -11.8 | 24.1 | 24.1 | 5.0 | 5.0 | 0.0 | 1.5 | 0.1 | 10.65 | 16.2 |
Standard lamp intensity analysis (0 sec, 1 rows)
| Serial | min | max | filter | fixable | 30-day history |
| #224 | 2296940 | 2315630 | 0 | yes |