Critical warnings from b-files' comments for days between 2023247 and 2023247 (0 sec, 2 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2023 246 05 40 53 Motor #2: appears stuck with a reference sensor activated.21#22405:32:041-1
rl:2023 247 11 04 59 New value set for watermark SUPPLY.VOLTAGE.LOW[93#22405:32:141-1

Instrument constant changes for the week before 2023247 (1 sec, 0 rows)

HG burned out? (AP results when HG is ON) (0 sec, 21 rows)

HP with step corrections larger than 20 steps (0 sec, 1 rows)

SerialYYYYJJJHP largemax correction
2242023247134.5

Non-critical warnings from b-files' comments for days between 2023247 and 2023247 (0 sec, 5 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2023 246 05 42 38 Bad command string. Command string buffer cleared.13#22405:32:071-1
rl:2023 246 05 42 38 Bad command string. Command string buffer cleared.13#22405:32:071-1
rl:2023 246 05 42 38 <13#22405:32:091-1
dp:Doing dispersion with internal HG.199#22410:05:351-1
ds:DS intensity too low ( 10/cy), skipping.199#22411:09:521-1

Measurements and tests summary.


_jday_loc_serial_skcdsdtedfmfzhghprssluvzs
247TORONTO#224calsc,Next schedule will be calsc (now calsc ZS= 90 ZS= 90.57599),Waiting until ZA =5722712727241418
(0 sec)

Number of days since last dome/window cleaning (0 sec, 2 rows)

serialdays
224more than a year

Missing files (0 sec, 1 rows)

Seriallast datadays ago
2242023244003

Questionable Run/Stop results. (0 sec, 1 rows)

SerialJdayslit 0dark rsdark countsslit 1slit 2slit 3slit 4slit 5slits 3+5
22420232470.9831.03628 / 29 (1.5M)0.9760.9750.9750.9760.9860.988

Run/Stop results for last week. (0 sec, 1 rows)

SerialTotalFailed 0Failed darkFailed 1Failed 2Failed 3Failed 4Failed 5Failed double
2241570777779

AP results for SB Brewers. (0 sec, 1 rows)

SerialJdayt1 offt1 ont2 offt2 ont3 offt3 onHV offHV on+12V off+12V on+5V off+5V on-12V off-12V on+24V off+24V on+5V sps off+5V sps onSL A offSL A onSL V offSL V onWatts
224202324731.131.132.132.131.031.01168.41168.012.012.05.25.2-12.0-11.924.124.15.05.00.01.50.110.5515.9

Standard lamp intensity analysis (0 sec, 1 rows)

Serialminmaxfilterfixable30-day history
#224232899023321500yes