Database last updated on Wed, 02 Sep 2026 02:11:10 +0000 by localhost

Critical warnings from b-files' comments for days between 2023245 and 2023245 (0 sec, 1 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2023 244 05 37 02 Motor #2: appears stuck with a reference sensor activated.31#22405:29:071-1

Instrument constant changes for the week before 2023245 (3 sec, 0 rows)

HG burned out? (AP results when HG is ON) (0 sec, 24 rows)

HP with step corrections larger than 20 steps (0 sec, 1 rows)

SerialYYYYJJJHP largemax correction
2242023245127.5

Non-critical warnings from b-files' comments for days between 2023245 and 2023245 (0 sec, 7 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2023 244 05 37 30 New value set for watermark SUPPLY.VOLTAGE.LOW[23#22405:29:101-1
rl:2023 244 05 38 39 Bad command string. Command string buffer cleared.13#22405:29:121-1
rl:2023 244 05 38 39 Bad command string. Command string buffer cleared.13#22405:29:121-1
rl:2023 244 05 38 39 <13#22405:29:141-1
dp:Doing dispersion with internal HG.299#22406:26:371-1
sc:SC intensity too low ( 359/cy), skipping.13#22420:15:291-1
ds:DS intensity too high ( 262027/cy) quitting.199#22420:45:151-1

Measurements and tests summary.


_jday_loc_serial_skcdsdtedfmhghprssluvzs
245TORONTO#224calsc,Next schedule will be calsc (now calsc ZS= 90 ZS= 90.59101)4722122325251530
(0 sec)

Number of days since last dome/window cleaning (0 sec, 2 rows)

serialdays
224more than a year

Missing files (0 sec, 1 rows)

Seriallast datadays ago
2242023244001

Questionable Run/Stop results. (0 sec, 1 rows)

SerialJdayslit 0dark rsdark countsslit 1slit 2slit 3slit 4slit 5slits 3+5
22420232450.9881.10028 / 28 (1.4M)0.9930.9920.9960.9980.9970.981

Run/Stop results for last week. (0 sec, 1 rows)

SerialTotalFailed 0Failed darkFailed 1Failed 2Failed 3Failed 4Failed 5Failed double
22415808888710

AP results for SB Brewers. (0 sec, 1 rows)

SerialJdayt1 offt1 ont2 offt2 ont3 offt3 onHV offHV on+12V off+12V on+5V off+5V on-12V off-12V on+24V off+24V on+5V sps off+5V sps onSL A offSL A onSL V offSL V onWatts
224202324521.421.422.122.221.421.41168.11168.412.012.05.25.2-11.9-11.924.124.15.05.00.01.50.010.6216.1

Standard lamp intensity analysis (0 sec, 1 rows)

Serialminmaxfilterfixable30-day history
#224206543023604900yes