Database last updated on Mon, 02 Feb 2026 21:10:53 +0000 by localhost

Critical warnings from b-files' comments for days between 2023237 and 2023237 (0 sec, 3 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2023 236 05 38 32 Motor #2: appears stuck with a reference sensor activated.21#22405:29:261-1
rl:2023 236 05 39 41 Command not accepted while another in progress.11#22405:29:291-1
rl:2023 237 16 05 23 New value set for watermark SUPPLY.VOLTAGE.LOW[53#22405:29:391-1

Instrument constant changes for the week before 2023237 (2 sec, 0 rows)

HG burned out? (AP results when HG is ON) (0 sec, 29 rows)

HP with step corrections larger than 20 steps (0 sec, 1 rows)

SerialYYYYJJJHP largemax correction
2242023237127.8

Non-critical warnings from b-files' comments for days between 2023237 and 2023237 (0 sec, 7 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
zs:ZS intensity too low ( 15/cy), skipping.299#22400:23:161-1
rl:2023 236 05 40 38 Bad command string. Command string buffer cleared.13#22405:29:321-1
rl:2023 236 05 40 38 Bad command string. Command string buffer cleared.13#22405:29:321-1
rl:2023 236 05 40 38 <13#22405:29:341-1
dp:Doing dispersion with internal HG.199#22409:54:551-1
ds:Suppressed oscillation of filter wheel.35#22415:26:401-1
ds:DS intensity too high ( 244012/cy) quitting.299#22415:44:361-1

Measurements and tests summary.


_jday_loc_serial_skcdsdtedfmhghprssluvzs
237TORONTO#224calsc,Next schedule will be calsc (now calsc ZS= 90 ZS= 91.473),Next schedule will be calsc (now calsc ZS= 90 ZS= 92.122),Waiting until ZA =612272425261628
(0 sec)

Number of days since last dome/window cleaning (0 sec, 2 rows)

serialdays
224more than a year

Missing files (0 sec, 1 rows)

Seriallast datadays ago
2242023236001

Questionable Run/Stop results. (0 sec, 1 rows)

SerialJdayslit 0dark rsdark countsslit 1slit 2slit 3slit 4slit 5slits 3+5
22420232370.9831.52926 / 17 (1.4M)0.9810.9700.9670.9810.9890.988

Run/Stop results for last week. (0 sec, 1 rows)

SerialTotalFailed 0Failed darkFailed 1Failed 2Failed 3Failed 4Failed 5Failed double
22416150121413151412

AP results for SB Brewers. (0 sec, 1 rows)

SerialJdayt1 offt1 ont2 offt2 ont3 offt3 onHV offHV on+12V off+12V on+5V off+5V on-12V off-12V on+24V off+24V on+5V sps off+5V sps onSL A offSL A onSL V offSL V onWatts
224202323723.123.124.024.023.323.31168.21168.912.012.05.25.2-12.0-12.023.923.95.15.10.01.50.010.3515.4

Standard lamp intensity analysis (0 sec, 1 rows)

Serialminmaxfilterfixable30-day history
#224199413020636100yes