Critical warnings from b-files' comments for days between 2023233 and 2023233 (0 sec, 2 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2023 232 05 40 36 Motor #2: appears stuck with a reference sensor activated.21#22405:29:271-1
rl:2023 233 02 56 49 New value set for watermark SUPPLY.VOLTAGE.LOW[53#22405:29:371-1

Instrument constant changes for the week before 2023233 (1 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 1 rows)

HG burned out? (AP results when HG is ON) (0 sec, 24 rows)

HP with step corrections larger than 20 steps (0 sec, 1 rows)

SerialYYYYJJJHP largemax correction
2242023233123.5

Non-critical warnings from b-files' comments for days between 2023233 and 2023233 (1 sec, 7 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
zs:ZS intensity too low ( 124/cy), skipping.399#22400:10:161-1
rl:2023 232 05 41 47 Bad command string. Command string buffer cleared.13#22405:29:291-1
rl:2023 232 05 41 47 Bad command string. Command string buffer cleared.13#22405:29:291-1
rl:2023 232 05 41 47 <13#22405:29:311-1
dp:Doing dispersion with internal HG.199#22410:02:431-1
sc:SC intensity too low ( 24/cy), skipping.23#22412:10:111-1
sc:Suppressed oscillation of filter wheel.15#22421:39:121-1

Measurements and tests summary.


_jday_loc_serial_skcdsdtedfzhghprssluvzs
233TORONTO#224calsc,Next schedule will be calsc (now calsc ZS= 90 ZS= 90.516),Next schedule will be calsc (now calsc ZS= 90 ZS= 91.163)632222425271720
(1 sec)

Number of days since last dome/window cleaning (0 sec, 2 rows)

serialdays
224more than a year

Questionable Run/Stop results. (0 sec, 1 rows)

SerialJdayslit 0dark rsdark countsslit 1slit 2slit 3slit 4slit 5slits 3+5
22420232330.9671.75032 / 26 (1.2M)0.9380.9360.9500.9420.9560.974

Run/Stop results for last week. (0 sec, 1 rows)

SerialTotalFailed 0Failed darkFailed 1Failed 2Failed 3Failed 4Failed 5Failed double
22416131121313131210

AP results for SB Brewers. (0 sec, 1 rows)

SerialJdayt1 offt1 ont2 offt2 ont3 offt3 onHV offHV on+12V off+12V on+5V off+5V on-12V off-12V on+24V off+24V on+5V sps off+5V sps onSL A offSL A onSL V offSL V onWatts
224202323324.424.425.425.424.424.41168.51168.112.012.05.25.2-12.0-12.023.823.85.05.00.01.50.010.4715.7

Standard lamp intensity analysis (0 sec, 1 rows)

Serialminmaxfilterfixable30-day history
#224201024021187200yes