Database last updated on Mon, 02 Feb 2026 06:10:57 +0000 by localhost

Critical warnings from b-files' comments for days between 2023229 and 2023229 (0 sec, 2 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2023 228 05 38 16 Motor #2: appears stuck with a reference sensor activated.21#22405:29:071-1
sc:SC intensity too low ( 276/cy), skipping.53#22412:23:171-1

Instrument constant changes for the week before 2023229 (2 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 1 rows)

HG burned out? (AP results when HG is ON) (0 sec, 27 rows)

HP with step corrections larger than 20 steps (0 sec, 1 rows)

SerialYYYYJJJHP largemax correction
2242023229127.3

Non-critical warnings from b-files' comments for days between 2023229 and 2023229 (0 sec, 7 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
zs:ZS intensity too low ( 2/cy), skipping.299#22400:52:511-1
rl:2023 228 05 39 27 Bad command string. Command string buffer cleared.13#22405:29:101-1
rl:2023 228 05 39 27 Bad command string. Command string buffer cleared.13#22405:29:101-1
rl:2023 228 05 39 27 <13#22405:29:121-1
rl:2023 229 06 59 34 New value set for watermark SUPPLY.VOLTAGE.LOW[33#22405:29:171-1
dp:Doing dispersion with internal HG.199#22409:53:581-1
ds:Suppressed oscillation of filter wheel.15#22419:51:591-1

Measurements and tests summary.


_jday_loc_serial_skcdsdtedfzhghprssluvzs
229TORONTO#224calsc,Next schedule will be calsc (now calsc ZS= 90 ZS= 90.519),Next schedule will be calsc (now calsc ZS= 90 ZS= 91.158)612222324261628
(0 sec)

Number of days since last dome/window cleaning (0 sec, 2 rows)

serialdays
224more than a year

Questionable Run/Stop results. (0 sec, 1 rows)

SerialJdayslit 0dark rsdark countsslit 1slit 2slit 3slit 4slit 5slits 3+5
22420232290.9791.47631 / 21 (1.4M)0.9870.9920.9910.9990.9850.995

Run/Stop results for last week. (0 sec, 1 rows)

SerialTotalFailed 0Failed darkFailed 1Failed 2Failed 3Failed 4Failed 5Failed double
2241682891181110

AP results for SB Brewers. (0 sec, 1 rows)

SerialJdayt1 offt1 ont2 offt2 ont3 offt3 onHV offHV on+12V off+12V on+5V off+5V on-12V off-12V on+24V off+24V on+5V sps off+5V sps onSL A offSL A onSL V offSL V onWatts
224202322923.123.123.923.923.123.11168.71168.412.012.05.25.2-11.9-11.923.923.95.05.00.01.50.110.4715.8

Standard lamp intensity analysis (0 sec, 1 rows)

Serialminmaxfilterfixable30-day history
#224199072020830200yes