Critical warnings from b-files' comments for days between 2023228 and 2023228 (0 sec, 2 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2023 227 05 40 56 Motor #2: appears stuck with a reference sensor activated.41#22405:29:011-1
rl:2023 227 05 40 58 New value set for watermark SUPPLY.VOLTAGE.HIGH[+5VSEC]: 5.357954 Volts.293#22405:29:041-1

Instrument constant changes for the week before 2023228 (1 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 1 rows)

HG burned out? (AP results when HG is ON) (0 sec, 29 rows)

HP with step corrections larger than 20 steps (0 sec, 1 rows)

SerialYYYYJJJHP largemax correction
2242023228128.5

Non-critical warnings from b-files' comments for days between 2023228 and 2023228 (1 sec, 12 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
ds:DS intensity too low ( 2/cy), skipping.299#22400:52:431-1
zs:ZS intensity too low ( 2/cy), skipping.299#22400:52:591-1
rl:2023 227 05 42 07 Bad command string. Command string buffer cleared.13#22405:29:231-1
rl:2023 227 05 42 07 Bad command string. Command string buffer cleared.13#22405:29:231-1
rl:2023 227 05 42 07 <13#22405:29:251-1
hg:HOME key pressed199#22413:58:591-1
pd:Going to menu199#22413:59:021-1
az:Azimuth discrepancy = 6501199#22414:05:246501-3030
si:New SI, HC =199#22414:15:001-1
ds:DS intensity too high ( 274756/cy) quitting.299#22415:25:121-1
ds:Suppressed oscillation of filter wheel.15#22417:51:041-1
sc:SC intensity too low ( 315/cy), skipping.13#22422:35:371-1

Measurements and tests summary.


_jday_loc_serial_skcdsdtedfzhghprssluvzs
228TORONTO#224calsc,Next schedule will be calsc (now calsc ZS= 90 ZS= 91.14499),Next schedule will be calsc (now calsc ZS= 90 ZS= 91.783)642212123261416
(1 sec)

Number of days since last dome/window cleaning (0 sec, 2 rows)

serialdays
224more than a year

Questionable Run/Stop results. (0 sec, 1 rows)

SerialJdayslit 0dark rsdark countsslit 1slit 2slit 3slit 4slit 5slits 3+5
22420232280.9841.40035 / 25 (1.4M)0.9640.9880.9920.9780.9750.990

Run/Stop results for last week. (0 sec, 1 rows)

SerialTotalFailed 0Failed darkFailed 1Failed 2Failed 3Failed 4Failed 5Failed double
22416826810899

AP results for SB Brewers. (0 sec, 1 rows)

SerialJdayt1 offt1 ont2 offt2 ont3 offt3 onHV offHV on+12V off+12V on+5V off+5V on-12V off-12V on+24V off+24V on+5V sps off+5V sps onSL A offSL A onSL V offSL V onWatts
224202322825.725.626.526.525.625.61168.71168.712.012.05.25.2-11.9-11.923.723.65.25.20.01.50.110.5315.9

Standard lamp intensity analysis (0 sec, 1 rows)

Serialminmaxfilterfixable30-day history
#224199810021646400yes