Database last updated on Wed, 04 Feb 2026 02:40:52 +0000 by localhost

Critical warnings from b-files' comments for days between 2023226 and 2023226 (0 sec, 3 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2023 225 05 40 05 Motor #2: appears stuck with a reference sensor activated.21#22405:29:371-1
rl:2023 225 05 40 19 Command not accepted while another in progress.11#22405:29:391-1
rl:2023 226 07 47 28 New value set for watermark SUPPLY.VOLTAGE.HIGH[243#22405:29:491-1

Instrument constant changes for the week before 2023226 (2 sec, 0 rows)

HG burned out? (AP results when HG is ON) (0 sec, 24 rows)

HP with step corrections larger than 20 steps (0 sec, 1 rows)

SerialYYYYJJJHP largemax correction
2242023226127.8

Non-critical warnings from b-files' comments for days between 2023226 and 2023226 (1 sec, 5 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2023 225 05 41 16 Bad command string. Command string buffer cleared.13#22405:29:421-1
rl:2023 225 05 41 16 Bad command string. Command string buffer cleared.13#22405:29:421-1
rl:2023 225 05 41 16 <13#22405:29:441-1
dp:Doing dispersion with internal HG.199#22410:02:071-1
sc:SC intensity too low ( 113/cy), skipping.13#22422:37:431-1

Measurements and tests summary.


_jday_loc_serial_skcdsdtedfzhghprssluvzs
226TORONTO#224calsc,Next schedule will be calsc (now calsc ZS= 90 ZS= 90.916),Next schedule will be calsc (now calsc ZS= 90 ZS= 91.551)572212223271621
(1 sec)

Number of days since last dome/window cleaning (0 sec, 2 rows)

serialdays
224more than a year

Questionable Run/Stop results. (0 sec, 1 rows)

SerialJdayslit 0dark rsdark countsslit 1slit 2slit 3slit 4slit 5slits 3+5
22420232260.9961.38129 / 21 (1.5M)0.9960.9970.9870.9940.9970.994

Run/Stop results for last week. (0 sec, 1 rows)

SerialTotalFailed 0Failed darkFailed 1Failed 2Failed 3Failed 4Failed 5Failed double
2241671478788

AP results for SB Brewers. (0 sec, 1 rows)

SerialJdayt1 offt1 ont2 offt2 ont3 offt3 onHV offHV on+12V off+12V on+5V off+5V on-12V off-12V on+24V off+24V on+5V sps off+5V sps onSL A offSL A onSL V offSL V onWatts
224202322623.223.223.723.723.123.11168.91168.412.012.05.25.2-11.8-11.823.723.75.15.10.01.50.010.6216.1

Standard lamp intensity analysis (0 sec, 1 rows)

Serialminmaxfilterfixable30-day history
#224227524023493900yes