Critical warnings from b-files' comments for days between 2023211 and 2023211 (1 sec, 3 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2023 210 05 48 43 Motor #2: appears stuck with a reference sensor activated.21#22405:42:071-1
rl:2023 210 06 48 22 New value set for watermark SUPPLY.VOLTAGE.HIGH[213#22405:42:171-1
sc:SC intensity too low ( 253/cy), skipping.53#22417:27:101-1

Instrument constant changes for the week before 2023211 (2 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 1 rows)

HG burned out? (AP results when HG is ON) (0 sec, 22 rows)

HP with step corrections larger than 20 steps (0 sec, 1 rows)

SerialYYYYJJJHP largemax correction
2242023211128.0

Non-critical warnings from b-files' comments for days between 2023211 and 2023211 (0 sec, 7 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
zs:ZS intensity too low ( 14/cy), skipping.399#22401:02:291-1
rl:2023 210 05 49 54 Bad command string. Command string buffer cleared.13#22405:42:101-1
rl:2023 210 05 49 54 Bad command string. Command string buffer cleared.13#22405:42:101-1
rl:2023 210 05 49 54 <13#22405:42:121-1
rl:2023 210 06 48 22 New value set for watermark SUPPLY.VOLTAGE.HIGH[13#22405:42:171-1
ds:DS intensity too high ( 624968/cy) quitting.399#22416:30:161-1
sc:Suppressed oscillation of filter wheel.15#22419:39:051-1

Measurements and tests summary.


_jday_loc_serial_skcdsdtedfmfzhghprssluvzs
211TORONTO#224calsc,Next schedule will be calsc (now calsc ZS= 90 ZS= 90.585),Next schedule will be calsc (now calsc ZS= 90 ZS= 91.194)6322822527251827
(0 sec)

Number of days since last dome/window cleaning (0 sec, 2 rows)

serialdays
224more than a year

Missing files (0 sec, 1 rows)

Seriallast datadays ago
2242023210001

Run/Stop results for last week. (0 sec, 1 rows)

SerialTotalFailed 0Failed darkFailed 1Failed 2Failed 3Failed 4Failed 5Failed double
2241601000000

AP results for SB Brewers. (0 sec, 1 rows)

SerialJdayt1 offt1 ont2 offt2 ont3 offt3 onHV offHV on+12V off+12V on+5V off+5V on-12V off-12V on+24V off+24V on+5V sps off+5V sps onSL A offSL A onSL V offSL V onWatts
224202321122.622.623.423.422.622.61168.71168.712.012.05.25.2-11.8-11.824.024.05.15.10.01.50.010.6216.1

Standard lamp intensity analysis (1 sec, 1 rows)

Serialminmaxfilterfixable30-day history
#224235138023519400yes