Critical warnings from b-files' comments for days between 2023210 and 2023210 (1 sec, 2 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| rl: | 2023 209 05 43 58 Motor #2: appears stuck with a reference sensor activated. | 3 | 1 | B21023.224 | TORON | #224 | 05:39:51 | | 1 | -1 |
| rl: | 2023 210 14 38 50 New value set for watermark SUPPLY.VOLTAGE.HIGH[ | 20 | 3 | B21023.224 | TORON | #224 | 05:40:04 | | 1 | -1 |
Instrument constant changes for the week before 2023210 (3 sec, 0 rows)
SL burned out? (AP results when SL is ON) (0 sec, 2 rows)
HG burned out? (AP results when HG is ON) (0 sec, 23 rows)
HP with step corrections larger than 20 steps (0 sec, 1 rows)
| Serial | YYYYJJJ | HP large | max correction |
| 224 | 2023210 | 2 | 30.4 |
Non-critical warnings from b-files' comments for days between 2023210 and 2023210 (0 sec, 10 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| zs: | ZS intensity too low ( 10/cy), skipping. | 3 | 99 | B21023.224 | TORON | #224 | 01:05:24 | | 1 | -1 |
| rl: | 2023 209 05 46 04 Bad command string. Command string buffer cleared. | 1 | 3 | B21023.224 | TORON | #224 | 05:39:54 | | 1 | -1 |
| rl: | 2023 209 05 46 04 Bad command string. Command string buffer cleared. | 1 | 3 | B21023.224 | TORON | #224 | 05:39:54 | | 1 | -1 |
| rl: | 2023 209 05 46 04 <x | 1 | 3 | B21023.224 | TORON | #224 | 05:39:56 | | 1 | -1 |
| dp: | Doing dispersion with internal HG. | 1 | 99 | B21023.224 | TORON | #224 | 09:55:42 | | 1 | -1 |
| sc: | SC intensity too low ( 31/cy), skipping. | 1 | 3 | B21023.224 | TORON | #224 | 11:37:31 | | 1 | -1 |
| hg: | HOME key pressed | 1 | 99 | B21023.224 | TORON | #224 | 18:49:46 | | 1 | -1 |
| pd: | Going to menu | 1 | 99 | B21023.224 | TORON | #224 | 18:49:49 | | 1 | -1 |
| ds: | DS intensity too high ( 359356/cy) quitting. | 3 | 99 | B21023.224 | TORON | #224 | 19:49:19 | | 1 | -1 |
| ds: | Suppressed oscillation of filter wheel. | 2 | 5 | B21023.224 | TORON | #224 | 21:45:47 | | 1 | -1 |
Measurements and tests summary.
| _jday | _loc | _serial | _skc | ds | dt | ed | fm | fz | hg | hp | rs | sl | uv | zs |
| 210 | TORONTO | #224 | 70 | 2 | 2 | 14 | 2 | 28 | 29 | 2 | 5 | 17 | 27 |
(1 sec)
Number of days since last dome/window cleaning (0 sec, 2 rows)
| serial | days |
| 224 | more than a year |
Run/Stop results for last week. (0 sec, 1 rows)
| Serial | Total | Failed 0 | Failed dark | Failed 1 | Failed 2 | Failed 3 | Failed 4 | Failed 5 | Failed double |
| 224 | 16 | 0 | 1 | 0 | 0 | 0 | 0 | 0 | 0 |
AP results for SB Brewers. (0 sec, 1 rows)
| Serial | Jday | t1 off | t1 on | t2 off | t2 on | t3 off | t3 on | HV off | HV on | +12V off | +12V on | +5V off | +5V on | -12V off | -12V on | +24V off | +24V on | +5V sps off | +5V sps on | SL A off | SL A on | SL V off | SL V on | Watts |
| 224 | 2023210 | 24.4 | 24.4 | 25.3 | 25.3 | 24.5 | 24.5 | 1168.7 | 1168.4 | 12.0 | 12.0 | 5.2 | 5.2 | -11.9 | -11.9 | 24.1 | 24.1 | 5.1 | 5.1 | 0.0 | 1.5 | 0.0 | 10.62 | 16.1 |
Standard lamp intensity analysis (0 sec, 1 rows)
| Serial | min | max | filter | fixable | 30-day history |
| #224 | 2324320 | 2325430 | 0 | yes |