Critical warnings from b-files' comments for days between 2023180 and 2023180 (0 sec, 1 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| rl: | 2023 179 05 35 38 Motor #2: appears stuck with a reference sensor activated. | 2 | 1 | B18023.224 | TORON | #224 | 05:29:52 | | 1 | -1 |
Instrument constant changes for the week before 2023180 (2 sec, 0 rows)
SL burned out? (AP results when SL is ON) (0 sec, 3 rows)
HG burned out? (AP results when HG is ON) (0 sec, 19 rows)
HP with step corrections larger than 20 steps (0 sec, 1 rows)
| Serial | YYYYJJJ | HP large | max correction |
| 224 | 2023180 | 1 | 27.7 |
Non-critical warnings from b-files' comments for days between 2023180 and 2023180 (0 sec, 8 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| zs: | ZS intensity too low ( 19/cy), skipping. | 3 | 99 | B18023.224 | TORON | #224 | 01:19:31 | | 1 | -1 |
| rl: | 2023 179 05 37 44 Bad command string. Command string buffer cleared. | 1 | 3 | B18023.224 | TORON | #224 | 05:29:54 | | 1 | -1 |
| rl: | 2023 179 05 37 44 Bad command string. Command string buffer cleared. | 1 | 3 | B18023.224 | TORON | #224 | 05:29:54 | | 1 | -1 |
| rl: | 2023 179 05 37 44 <# | 1 | 3 | B18023.224 | TORON | #224 | 05:29:56 | | 1 | -1 |
| rl: | 2023 179 05 42 53 New value set for watermark SUPPLY.VOLTAGE.HIGH[ | 1 | 3 | B18023.224 | TORON | #224 | 05:30:01 | | 1 | -1 |
| rl: | 2023 180 07 09 55 New value set for watermark SUPPLY.VOLTAGE.HIGH[ | 2 | 3 | B18023.224 | TORON | #224 | 05:30:04 | | 1 | -1 |
| ds: | DS intensity too low ( 6/cy), skipping. | 3 | 99 | B18023.224 | TORON | #224 | 10:09:15 | | 1 | -1 |
| ds: | DS intensity too high ( 211962/cy) quitting. | 1 | 99 | B18023.224 | TORON | #224 | 18:24:50 | | 1 | -1 |
Measurements and tests summary.
| _jday | _loc | _serial | _skc | ds | dt | ed | fm | fz | hg | hp | rs | sl | uv | zs |
| 180 | TORONTO | #224 | calsc,Next schedule will be calsc (now calsc ZS= 90 ZS= 91.233),Next schedule will be calsc (now calsc ZS= 90 ZS= 91.801) | 76 | 2 | 2 | 16 | 1 | 25 | 26 | 2 | 5 | 18 | 14 |
(0 sec)
Number of days since last dome/window cleaning (0 sec, 2 rows)
| serial | days |
| 224 | more than a year |
Missing files (0 sec, 1 rows)
| Serial | last data | days ago |
| 224 | 2023181 | 0-1 |
Run/Stop results for last week. (0 sec, 1 rows)
| Serial | Total | Failed 0 | Failed dark | Failed 1 | Failed 2 | Failed 3 | Failed 4 | Failed 5 | Failed double |
| 224 | 16 | 0 | 1 | 0 | 0 | 0 | 0 | 0 | 0 |
AP results for SB Brewers. (0 sec, 1 rows)
| Serial | Jday | t1 off | t1 on | t2 off | t2 on | t3 off | t3 on | HV off | HV on | +12V off | +12V on | +5V off | +5V on | -12V off | -12V on | +24V off | +24V on | +5V sps off | +5V sps on | SL A off | SL A on | SL V off | SL V on | Watts |
| 224 | 2023180 | 22.6 | 22.6 | 23.7 | 23.7 | 22.3 | 22.4 | 1167.8 | 1168.1 | 12.0 | 12.0 | 5.2 | 5.2 | -11.9 | -11.9 | 24.1 | 24.1 | 5.0 | 5.0 | 0.0 | 1.5 | 0.1 | 10.60 | 16.1 |
Standard lamp intensity analysis (0 sec, 1 rows)
| Serial | min | max | filter | fixable | 30-day history |
| #224 | 2351320 | 2352560 | 0 | yes |