Database last updated on Thu, 03 Sep 2026 12:11:11 +0000 by localhost

Critical warnings from b-files' comments for days between 2023131 and 2023131 (0 sec, 2 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2023 130 05 34 25 New value set for watermark SUPPLY.VOLTAGE.HIGH[973B13123.224TORON#22405:20:341-1
rl:2023 130 05 36 01 Motor #2: appears stuck with a reference sensor activated.11B13123.224TORON#22405:21:141-1

Instrument constant changes for the week before 2023131 (3 sec, 0 rows)

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
2242023131315.4

Non-critical warnings from b-files' comments for days between 2023131 and 2023131 (0 sec, 10 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
zs:ZS intensity too low ( 8/cy), skipping.399B13123.224TORON#22400:56:071-1
rl:Log item overwritten by later entry.13B13123.224TORON#22405:20:311-1
rl:2023 130 05 34 50 Bad command string. Command string buffer cleared.13B13123.224TORON#22405:20:561-1
rl:2023 130 05 34 50 Bad command string. Command string buffer cleared.13B13123.224TORON#22405:20:561-1
rl:2023 130 05 34 50 < 13B13123.224TORON#22405:20:581-1
dp:Doing dispersion with internal HG.199B13123.224TORON#22409:48:581-1
sc:SC intensity too low ( 123/cy), skipping.13B13123.224TORON#22411:30:061-1
checkhg:Last HG had to adjust the micrometer by at least 1199B13123.224TORON#22414:20:101-1
dsp:Doing dispersion test with Hg (UV port)199B13123.224TORON#22414:23:071-1
ds:DS intensity too high ( 299958/cy) quitting.199B13123.224TORON#22420:36:461-1

Measurements and tests summary.


_jday_loc_serial_skcdsdtedfzhghprssluvzs
131TORONTO#224menu,Next schedule will be calsc (now calsc ZS= 90 ZS= 90.607),Next schedule will be calsc (now calsc ZS= 90 ZS= 91.182)432222724251014
(0 sec)

Number of days since last dome/window cleaning (1 sec, 2 rows)

serialdays
224more than a year

Run/Stop results for last week. (0 sec, 1 rows)

SerialTotalFailed 0Failed darkFailed 1Failed 2Failed 3Failed 4Failed 5Failed double
2241600000000

AP results for SB Brewers. (0 sec, 1 rows)

SerialJdayt1 offt1 ont2 offt2 ont3 offt3 onHV offHV on+12V off+12V on+5V off+5V on-12V off-12V on+24V off+24V on+5V sps off+5V sps onSL A offSL A onSL V offSL V onWatts
224202313124.424.525.425.523.923.91168.11167.814.514.55.75.7-12.9-12.830.030.05.85.80.01.50.010.5516.0

Standard lamp intensity analysis (0 sec, 1 rows)

Serialminmaxfilterfixable30-day history
#224283084028313600yes