Critical warnings from b-files' comments for days between 2021009 and 2021009 (0 sec, 1 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2021 008 05 34 42 Motor #2: appears stuck with a reference sensor activated.21B00921.224TORON#22405:33:291-1

Instrument constant changes for the week before 2021009 (1 sec, 0 rows)

Non-critical warnings from b-files' comments for days between 2021009 and 2021009 (0 sec, 9 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2021 008 05 33 32 Reset Warm -> Tepid from 2021 008 05 33 26 until 2021 008 05 33 27 . Counts: 1,013B00921.224TORON#22405:33:261-1
rl:2021 008 05 48 22 <M,6,242>13B00921.224TORON#22405:33:381-1
rl:2021 008 05 48 22 <M,6,242>15B00921.224TORON#22405:33:381-1
rl:2021 008 06 00 06 New value set for watermark LOOP.RUN.TIME: 699.000025 millisec.33B00921.224TORON#22405:33:401-1
rl:2021 008 06 00 06 New value set for watermark LOOP.RUN.TIME: 699.000025 millisec.33B00921.224TORON#22405:33:401-1
rl:2021 009 06 58 43 New value set for watermark LOOP.RUN.TIME: 721.000004 millisec.13B00921.224TORON#22405:33:471-1
rl:2021 009 06 58 43 New value set for watermark LOOP.RUN.TIME: 721.000004 millisec.13B00921.224TORON#22405:33:471-1
az:Azimuth discrepancy = -235199B00921.224TORON#22405:36:26-235-3030
zs:ZS intensity too low ( 486/cy), skipping.399B00921.224TORON#22413:10:001-1

Measurements and tests summary.


_jday_loc_serial_skcdsdtedfzhghprssluvzs
009TORONTO#224ecsc,Next schedule will be ecsc (now ecsc ZS= 93 ZS= 94.016)402131212273515
(1 sec)

Number of days since last dome/window cleaning (0 sec, 2 rows)

serialdays
224more than a year

Missing files (0 sec, 1 rows)

Seriallast datadays ago
22420210100-1

Run/Stop results for last week. (0 sec, 1 rows)

SerialTotalFailed 0Failed darkFailed 1Failed 2Failed 3Failed 4Failed 5Failed double
2241506000000

AP results for SB Brewers. (0 sec, 1 rows)

SerialJdayt1 offt1 ont2 offt2 ont3 offt3 onHV offHV on+12V off+12V on+5V off+5V on-12V off-12V on+24V off+24V on+5V sps off+5V sps onSL A offSL A onSL V offSL V onWatts
22420210096.36.37.07.05.05.01216.11216.112.012.05.15.1-12.1-12.124.723.95.05.00.01.50.09.9014.4

Standard lamp intensity analysis (0 sec, 1 rows)

Serialminmaxfilterfixable30-day history
#224451378045148600yes