Critical warnings from b-files' comments for days between 2020011 and 2020011 (0 sec, 1 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| rl: | 2020 010 05 39 14 Motor #2: appears stuck with a reference sensor activated. | 1 | 1 | B01120.224 | TORON | #224 | 05:37:13 | | 1 | -1 |
Instrument constant changes for the week before 2020011 (3 sec, 0 rows)
SL burned out? (AP results when SL is ON) (0 sec, 23 rows)
| Serial | YYYYJJJ | SL Current | SL Voltage |
| 224 | 2020004 | -12.11 | 23.84 |
| 224 | 2020005 | -12.1 | 23.84 |
| 224 | 2020006 | -12.1 | 23.84 |
| 224 | 2020007 | -12.1 | 23.84 |
| 224 | 2020008 | -12.1 | 23.84 |
| 224 | 2020009 | -12.08 | 23.84 |
| 224 | 2020010 | -12.11 | 23.84 |
| 224 | 2020011 | -12.11 | 23.84 |
HG burned out? (AP results when HG is ON) (0 sec, 10 rows)
Non-critical warnings from b-files' comments for days between 2020011 and 2020011 (0 sec, 3 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| rl: | 2020 010 05 50 46 <M,6,242> | 1 | 3 | B01120.224 | TORON | #224 | 05:37:20 | | 1 | -1 |
| rl: | 2020 010 05 50 46 <M,6,242> | 1 | 5 | B01120.224 | TORON | #224 | 05:37:20 | | 1 | -1 |
| re: | Illegal command (re) in schedule ecsc, ignoring | 1 | 99 | B01120.224 | TORON | #224 | 05:38:07 | | 1 | -1 |
Measurements and tests summary.
| _jday | _loc | _serial | _skc | ds | dt | ed | fm | fz | hg | hp | rs | sl | zs |
| 011 | TORONTO | #224 | ecsc,Next schedule will be ecsc (now ecsc ZS= 93 ZS= 93.70099) | 15 | 2 | 1 | 47 | 7 | 14 | 13 | 2 | 5 | 74 |
(0 sec)
Number of days since last dome/window cleaning (0 sec, 2 rows)
| serial | days |
| 224 | more than a year |
Missing files (0 sec, 1 rows)
| Serial | last data | days ago |
| 224 | 2020012 | 0-1 |
Run/Stop results for last week. (0 sec, 1 rows)
| Serial | Total | Failed 0 | Failed dark | Failed 1 | Failed 2 | Failed 3 | Failed 4 | Failed 5 | Failed double |
| 224 | 16 | 0 | 5 | 0 | 0 | 0 | 0 | 0 | 0 |
AP results for MB Brewer. (0 sec, 1 rows)
| Serial | Jday | t1 off | t1 on | t2 off | t2 on | t3 off | t3 on | HV off | HV on | +15V off | +15V on | +5V off | +5V on | -15V off | -15V on | +24V off | +24V on | +5V sps off | +5V sps on | -8V sps off | -8V sps on | SL A off | SL A on | SL V off | SL V on | W |
| 224 | 2020011 | 11.6 | 0.0 | 11.5 | 0.0 | 10.7 | 1.3 | 1348.7 | 2.2 | 12.0 | 7.5 | 5.1 | 0.8 | -12.1 | 2.0 | 24.6 | 2.2 | 5.0 | 12.0 | -5.0 | 5.1 | 0.0 | -12.1 | 0.0 | 23.84 | -288.7 |
Standard lamp intensity analysis (0 sec, 1 rows)
| Serial | min | max | filter | fixable | 30-day history |
| #224 | 3209840 | 3212610 | 0 | yes |