Database last updated on Wed, 09 Sep 2026 08:41:02 +0000 by localhost

Critical warnings from b-files' comments for days between 2020011 and 2020011 (0 sec, 1 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2020 010 05 39 14 Motor #2: appears stuck with a reference sensor activated.11B01120.224TORON#22405:37:131-1

Instrument constant changes for the week before 2020011 (3 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 23 rows)

SerialYYYYJJJSL CurrentSL Voltage
2242020004-12.1123.84
2242020005-12.123.84
2242020006-12.123.84
2242020007-12.123.84
2242020008-12.123.84
2242020009-12.0823.84
2242020010-12.1123.84
2242020011-12.1123.84

HG burned out? (AP results when HG is ON) (0 sec, 10 rows)

Non-critical warnings from b-files' comments for days between 2020011 and 2020011 (0 sec, 3 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2020 010 05 50 46 <M,6,242>13B01120.224TORON#22405:37:201-1
rl:2020 010 05 50 46 <M,6,242>15B01120.224TORON#22405:37:201-1
re:Illegal command (re) in schedule ecsc, ignoring199B01120.224TORON#22405:38:071-1

Measurements and tests summary.


_jday_loc_serial_skcdsdtedfmfzhghprsslzs
011TORONTO#224ecsc,Next schedule will be ecsc (now ecsc ZS= 93 ZS= 93.70099)152147714132574
(0 sec)

Number of days since last dome/window cleaning (0 sec, 2 rows)

serialdays
224more than a year

Missing files (0 sec, 1 rows)

Seriallast datadays ago
22420200120-1

Run/Stop results for last week. (0 sec, 1 rows)

SerialTotalFailed 0Failed darkFailed 1Failed 2Failed 3Failed 4Failed 5Failed double
2241605000000

AP results for MB Brewer. (0 sec, 1 rows)

SerialJdayt1 offt1 ont2 offt2 ont3 offt3 onHV offHV on+15V off+15V on+5V off+5V on-15V off-15V on+24V off+24V on+5V sps off+5V sps on-8V sps off-8V sps onSL A offSL A onSL V offSL V onW
224202001111.60.011.50.010.71.31348.72.212.07.55.10.8-12.12.024.62.25.012.0-5.05.10.0-12.10.023.84-288.7

Standard lamp intensity analysis (0 sec, 1 rows)

Serialminmaxfilterfixable30-day history
#224320984032126100yes