Database last updated on Wed, 09 Sep 2026 20:11:06 +0000 by localhost

Critical warnings from b-files' comments for days between 2020001 and 2020001 (0 sec, 1 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2020 365 05 33 23 Motor #2: appears stuck with a reference sensor activated.11B00120.224TORON#22405:28:341-1

Instrument constant changes for the week before 2020001 (2 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 29 rows)

SerialYYYYJJJSL CurrentSL Voltage
2242019359-12.123.84
2242019360-12.1123.81
2242019361-12.1123.81
2242019362-12.1123.81
2242019363-12.1123.81
2242019364-12.123.81
2242019365-12.1123.81
2242020001-12.123.81

Non-critical warnings from b-files' comments for days between 2020001 and 2020001 (1 sec, 3 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2020 365 05 44 52 <M,6,242>13B00120.224TORON#22405:28:401-1
rl:2020 365 05 44 52 <M,6,242>15B00120.224TORON#22405:28:401-1
re:Illegal command (re) in schedule ecsc, ignoring199B00120.224TORON#22405:32:271-1

Measurements and tests summary.


_jday_loc_serial_skcdsdtedfzhghprsslzs
001TORONTO#224ecsc,Next schedule will be ecsc (now ecsc ZS= 93 ZS= 93.06301)27211111102764
(1 sec)

Number of days since last dome/window cleaning (0 sec, 2 rows)

serialdays
224more than a year

Missing files (0 sec, 1 rows)

Seriallast datadays ago
22420200020-1

Run/Stop results for last week. (0 sec, 1 rows)

SerialTotalFailed 0Failed darkFailed 1Failed 2Failed 3Failed 4Failed 5Failed double
2241604000000

AP results for MB Brewer. (0 sec, 1 rows)

SerialJdayt1 offt1 ont2 offt2 ont3 offt3 onHV offHV on+15V off+15V on+5V off+5V on-15V off-15V on+24V off+24V on+5V sps off+5V sps on-8V sps off-8V sps onSL A offSL A onSL V offSL V onW
22420200016.80.07.20.05.91.31349.42.212.06.85.10.5-12.12.024.62.25.012.0-5.05.10.0-12.10.023.82-288.1

Standard lamp intensity analysis (0 sec, 1 rows)

Serialminmaxfilterfixable30-day history
#224320903032128900yes