Database last updated on Wed, 09 Sep 2026 19:40:58 +0000 by localhost

Critical warnings from b-files' comments for days between 2019327 and 2019327 (0 sec, 1 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2019 326 05 16 38 Motor #2: appears stuck with a reference sensor activated.11B32719.224TORON#22405:09:511-1

Instrument constant changes for the week before 2019327 (3 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 32 rows)

SerialYYYYJJJSL CurrentSL Voltage
2242019320-12.0823.84
2242019321-12.0723.84
2242019322-12.123.84
2242019323-12.123.81
2242019324-12.123.79
2242019325-12.1123.81
2242019326-12.1123.84
2242019327-12.123.81

Non-critical warnings from b-files' comments for days between 2019327 and 2019327 (0 sec, 4 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2019 326 05 28 08 <M,6,242>13B32719.224TORON#22405:09:571-1
rl:2019 326 05 28 08 <M,6,242>15B32719.224TORON#22405:09:571-1
re:Illegal command (re) in schedule ecsc, ignoring199B32719.224TORON#22405:12:221-1
fz:Suppressed oscillation of filter wheel.15B32719.224TORON#22413:17:141-1

Measurements and tests summary.


_jday_loc_serial_skcdsdtedfzhghprsslzs
327TORONTO#224ecsc,Next schedule will be ecsc (now ecsc ZS= 93 ZS= 93.009)6821816152637
(0 sec)

Number of days since last dome/window cleaning (0 sec, 2 rows)

serialdays
224346

Missing files (0 sec, 1 rows)

Seriallast datadays ago
22420193280-1

Run/Stop results for last week. (0 sec, 1 rows)

SerialTotalFailed 0Failed darkFailed 1Failed 2Failed 3Failed 4Failed 5Failed double
2241506000000

AP results for MB Brewer. (0 sec, 1 rows)

SerialJdayt1 offt1 ont2 offt2 ont3 offt3 onHV offHV on+15V off+15V on+5V off+5V on-15V off-15V on+24V off+24V on+5V sps off+5V sps on-8V sps off-8V sps onSL A offSL A onSL V offSL V onW
22420193277.90.08.20.07.21.31349.02.212.06.15.10.5-12.12.024.62.25.012.0-5.05.10.0-12.10.023.81-288.1

Standard lamp intensity analysis (0 sec, 1 rows)

Serialminmaxfilterfixable30-day history
#224315382031548300yes