Database last updated on Wed, 09 Sep 2026 21:40:58 +0000 by localhost

Critical warnings from b-files' comments for days between 2019320 and 2019320 (0 sec, 1 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2019 319 05 12 27 Motor #2: appears stuck with a reference sensor activated.21B32019.224TORON#22405:11:161-1

Instrument constant changes for the week before 2019320 (2 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 32 rows)

SerialYYYYJJJSL CurrentSL Voltage
2242019313-12.0823.81
2242019314-12.0823.81
2242019315-12.0823.81
2242019316-12.0823.84
2242019317-12.0723.84
2242019318-12.0823.84
2242019319-12.0823.84
2242019320-12.0823.84

HG burned out? (AP results when HG is ON) (0 sec, 7 rows)

Non-critical warnings from b-files' comments for days between 2019320 and 2019320 (0 sec, 8 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2019 319 05 12 02 Reset Warm -> Tepid from 2019 319 05 11 56 until 2019 319 05 11 57 . Counts: 1,013B32019.224TORON#22405:11:141-1
rl:2019 319 05 13 21 <M,6,242>23B32019.224TORON#22405:11:201-1
rl:2019 319 05 13 21 <M,6,242>25B32019.224TORON#22405:11:201-1
rl:2019 319 05 30 12 New value set for watermark LOOP.RUN.TIME: 674.000025 millisec.33B32019.224TORON#22405:11:311-1
rl:2019 319 05 30 12 New value set for watermark LOOP.RUN.TIME: 674.000025 millisec.33B32019.224TORON#22405:11:311-1
ds:DS intensity too high ( 215302/cy) quitting.399B32019.224TORON#22414:56:581-1
ds:Suppressed oscillation of filter wheel.15B32019.224TORON#22414:57:131-1
mf:Illegal command (mf) in schedule ecsc, ignoring499B32019.224TORON#22422:21:251-1

Measurements and tests summary.


_jday_loc_serial_skcdsdtedfmfzhghprsslzs
320TORONTO#224ecsc,Next schedule will be ecsc (now ecsc ZS= 93 ZS= 93.042)7421321011102737
(0 sec)

New operators? (0 sec, 28 rows)

Number of days since last dome/window cleaning (0 sec, 2 rows)

serialdays
224339

Missing files (0 sec, 1 rows)

Seriallast datadays ago
22420193210-1

Run/Stop results for last week. (0 sec, 1 rows)

SerialTotalFailed 0Failed darkFailed 1Failed 2Failed 3Failed 4Failed 5Failed double
2241606000000

AP results for MB Brewer. (0 sec, 1 rows)

SerialJdayt1 offt1 ont2 offt2 ont3 offt3 onHV offHV on+15V off+15V on+5V off+5V on-15V off-15V on+24V off+24V on+5V sps off+5V sps on-8V sps off-8V sps onSL A offSL A onSL V offSL V onW
22420193204.70.06.50.0-0.81.31349.42.112.05.45.10.4-12.12.024.62.25.012.0-5.05.10.0-12.10.023.81-288.0

Standard lamp intensity analysis (0 sec, 1 rows)

Serialminmaxfilterfixable30-day history
#224315346031591300yes